NICOLETONE VEEG 44 CHANNEL ETHERNET DIGITAL VI
Report
- Report Number
- 3008289288-2010-00004
- Event Type
- Other
- Date Received
- March 7, 2012
- Date of Event
- February 2, 2010
- Report Date
- March 3, 2010
- Manufacturer
- CAREFUSION 209 INC.
- Product Code
- GWQ
- PMA / PMN Number
- K991054
- Removal / Correction Number
- 2126317/03/02/2010/001-C
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
FOR UNITS IN THE FIELD, WE WILL NOTIFY USERS ABOUT THE R/L PROTOCOLS AND INSTRUCT THEM TO BE REMOVED FROM THE PROTOCOLS AND INSTRUCT THEM TO BE REMOVED FROM THE PROTOCOL MENU IF THEY ARE NOT UTILIZED AND PROPERLY UNDERSTAND BY THE USER FACILITY AND STAFF. ONE PT UNDER-WENT BILATERAL ELECTRODE GRID PLACEMENT WHEN UNILATERAL GRID PLACEMENT MAY HAVE SUFFICED. THE USER FACILITY ALSO SUBMITTED A MEDWATCH REPORT (B)(4). THIS REPORT WAS EDITED ON 3/6/2012 TO REVISE THE REGISTRATION NUMBER FROM 2126317 TO 3008289288. THE CURRENT REGISTRATION NUMBER WAS NOT KNOWN AT THE TIME THE INITIAL MDR WAS SENT.
A CUSTOMER COMPLAINT WAS RECEIVED INDICATING THAT THE USER HAD DISCOVERED THAT EEG DATA FROM ELECTRODES PLACED ON THE RIGHT SIDE OF THE HEAD WAS DISPLAYED IN THE EEG RECORD AS HAVING COME FROM THE LEFT SIDE OF THE HEAD AND VISA VERSA. INVESTIGATION DETERMINED THAT THE CUSTOMER SELECTION AND USE OF THE "ROUTINE V4 R/L" PROTOCOL (R/L REFERS TO THE RIGHT SIDE ELECTRODES DISPLAYED OVER THE LEFT SIDE ELECTRODES) WITHOUT THE ASSOCIATED AMPLIFIER OVERLAY MAY RESULT IN A MISMATCH BETWEEN THE AMPLIFIER LABELING AND THE PROTOCOL. THIS MAY RESULT IN THE MISINTERPRETATION OF THE LEFT VERSUS THE RIGHT SIDE OF THE BRAIN. IN ONE CASE, A PT UNDERWENT BILATERAL GRID PLACEMENT SURGERY WHEN UNILATERAL GRID PLACEMENT WOULD HAVE SUFFICED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NICOLETONE VEEG 44 CHANNEL ETHERNET DIGITAL VI | NICOLETONE | GWQ | CAREFUSION 209 INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other | EVALUATED ON CUSTOMER SITE. |