FDA Adverse Event Injury Summary report: N

DURASEAL, UNKNOWN

MDR report key: 24845384 · Received April 10, 2026

Report

Report Number
3003418325-2026-00010
Event Type
Injury
Date Received
April 10, 2026
Report Date
May 29, 2026
Manufacturer
INTEGRA - PRINCETON
Product Code
NQR
PMA / PMN Number
P040034
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

JOURNAL OF NEUROSURGERY PUBLISHED AN ARTICLE TITLED: ¿DURAL SEALANT-OMA¿ AFTER POSTERIOR FOSSA SURGERY: ILLUSTRATIVE CASE". BACKGROUND: DURAL SEALANTS, WIDELY USED IN NEUROSURGICAL PRACTICE, HAVE BEEN ASSOCIATED WITH FEW COMPLICATIONS, MOSTLY RELATING TO THE EXPANSIVE NATURE OF THE HYDROGEL. ALTHOUGH GENERALLY WELL TOLERATED, DURASEAL EXPANDS UP TO 150% AFTER APPLICATION, AND RARE REPORTS HAVE DESCRIBED COMPLICATIONS FROM MASS EFFECT. ADVERSE EVENT: THIS REPORT PRESENTS THE CASE OF A 34-YEAR-OLD FEMALE WITH A PAST SURGICAL HISTORY OF MULTIPLE OPERATIONS FOR A CHIARI TYPE 1 MALFORMATION. AFTER A THOROUGH WORKUP, INTRAOPERATIVE IMAGES AND HISTOLOGICAL FINDINGS REVEALED A CONCEALED HEMATOMA-LIKE MASS AT THE PATIENT¿S PREVIOUS SURGICAL SITE, LIKELY DUE TO THE DURASEAL SYSTEM. IMAGING REVEALED THAT THE DURASEAL WAS MOSTLY HYPERINTENSE ON T2-WEIGHTED MRI WITH FOCAL SPOTS OF HYPO-/ISOINTENSITY, SIMILAR IN APPEARANCE TO AN ABSCESS OR HEMATOMA. HISTOLOGICAL ANALYSIS SHOWED EXUBERANT FIBROSIS, GRANULATION TISSUE, AND A GRANULOMATOUS RESPONSE TO THE FOREIGN MATERIAL USED FOR THE DURAL PATCH AND SEAL RATHER THAN THE WALL OF AN INFECTIOUS ABSCESS. CONCLUSION: THIS CASE ADDS TO THE GROWING BODY OF LITERATURE ON DURASEAL, HIGHLIGHTING THE IMAGING AND HISTOLOGICAL CHARACTERISTICS OF ¿DURAL SEALANT-OMAS¿ AND THEIR EXPANSILE NATURE THAT CAN LEAD TO THE FORMATION OF A MASS MIMICKING AN ABSCESS OR A HEMATOMA. IN AREAS SUSCEPTIBLE TO COMPRESSION, CLINICIANS OUGHT TO BE VIGILANT TO IDENTIFY FINDINGS CONSISTENT WITH THE APPEARANCE OF THESE DURAL SEALANT-OMAS. HTTPS://THEJNS.ORG/DOI/ABS/10.3171/CASE25574.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
34561 DURASEAL, UNKNOWN DURAL REPAIR NQR INTEGRA - PRINCETON

Patients

Seq Age Sex Outcome Treatment
1 34 YR Female Other