DENTAL SCREW
Report
- Report Number
- 0001038806-2026-01949
- Event Type
- Malfunction
- Date Received
- April 10, 2026
- Report Date
- April 16, 2026
- Manufacturer
- ZIMVIE US CORP LLC
- Product Code
- NHA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODES WERE ADDED: 4111. H6: INVESTIGATION FINDINGS CODE WAS ADDED: 3252. H6: INVESTIGATION CONCLUSIONS CODE WAS ADDED: 4307. H11: NARRATIVE/DATA WAS UPDATED. ZIMVIE RECEIVED A PORTION/FRAGMENT OF THE UNKNOWN SCREW FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICES IDENTIFIED THE FRACTURED SCREW FRAGMENT STUCK INSIDE THE IMPLANT. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. DHR, STERILIZATION, AND COMPLAINT HISTORY REVIEW COULD NOT BE PERFORMED, AS THE SUBJECT LOT NUMBER ASSOCIATED WITH THE UNKNOWN SCREW IS NOT AVAILABLE. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO PREVENT THE DISTRIBUTION OF NON-CONFORMING PRODUCT AND ENSURE THE PRODUCT IS WITHIN SPECIFICATIONS. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE EXCESSIVE OCCLUSAL FORCES - PATIENT FACTORS, OR CUSTOMER ERROR. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.
ZIMVIE COMPLAINT NUMBER (B)(4). G4: PMA/510(K) NUMBER NOT AVAILABLE. PRODUCT HAS BEEN RECEIVED BY ZIMVIE AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.
DURING PRODUCT EVALUATION A FRACTURED SCREW WAS IDENTIFIED INSIDE OF THE IMPLANT. FOLLOW UP WITH THE CUSTOMER REVEALED THAT THE CUSTOMER WAS NOT AWARE OF THE FRACTURED SCREW.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 34563 | DENTAL SCREW | NHA | ZIMVIE US CORP LLC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | Unknown | TSX37B10, TSX IMPLANT, 3.7MMD, 10MML |