UNICEL¿ DXL800 ACCESS® IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2012-00598
- Event Type
- Malfunction
- Date Received
- March 8, 2012
- Date of Event
- February 15, 2012
- Report Date
- February 16, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- LFX
- PMA / PMN Number
- K023764
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SERVICE WAS NOT DISPATCHED FOR THIS EVENT. BECKMAN COULTER INC. ASSESSMENT OF CUSTOMER SUPPLIED PERFORMANCE DATA INDICATED THAT A PREVIOUSLY EXECUTED ((B)(4) 2012) AND ACCEPTED CALIBRATION'S S0 CALIBRATOR RELATIVE LIGHT UNITS (RLUS) WERE HIGHER THAN BECKMAN COULTER INC. IN HOUSE RELEASE, AS WELL AS THE RLUS ASSOCIATED WITH THE (B)(4) 2012, QC RESULTS. BECKMAN COULTER INC. ASSESSMENT OF CUSTOMER SUPPLIED INSTRUMENT ASSAY PERFORMANCE DATA INDICATED THAT QUALITY CONTROL ASSESSMENTS WERE PERFORMED ON EACH DAY DURING THE TIMEFRAME OF THE EVENT. ALL QC VALUES WERE WITHIN TWO STANDARD DEVIATIONS OF THE CUSTOMER'S ESTABLISHED MEANS. A ROUTINE SYSTEM CHECK PERFORMED PRIOR TO THE EVENT PASSED WITHIN THE INSTRUMENT'S SPECIFICATIONS. THE CAUSE OF THIS EVENT IS UNKNOWN AND COULD NOT BE DETERMINED BASED ON THE SUPPLIED INFORMATION ASSOCIATED MDRS: 2122870-2012-00597, 2122870-2012-00598.
THE CUSTOMER REPORTED THAT NO VALUE RUBELLA IMMUNOGLOBULIN G (IGG) LOW LEVEL QUALITY CONTROL RESULTS WITH INSTRUMENT GENERATED FLAGS WERE GENERATED ON AN UNICEL DXL800 ACCESS IMMUNOASSAY SYSTEM ON TWO DAYS. THIS REPORT REPRESENTS THE NO VALUE RUBELLA IMMUNOGLOBULIN G (IGG) QUALITY CONTROL RESULTS GENERATED ON (B)(6) 2012. THE INSTRUMENT GENERATED FLAG WAS AN INDETERMINATE FLAG WHICH IS INDICATIVE OF A RESULT THAT CANNOT BE DISTINGUISHED FROM A SYSTEM FAILURE. THERE WERE NO PATIENT RESULTS GENERATED IN CONNECTION TO THIS EVENT. THERE HAVE BEEN NO REPORTS OF DEATH, SERIOUS INJURY OR MODIFICATION TO PATIENT TREATMENT ASSOCIATED OR ATTRIBUTED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL¿ DXL800 ACCESS® IMMUNOASSAY SYSTEM | ANALYZER, CHEMISTRY | LFX | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ACCESS RUBELLA IGG REAGENT AND CALIBRATOR |