ANALYTICAL E MODULE
Report
- Report Number
- 1823260-2012-01325
- Event Type
- Malfunction
- Date Received
- March 8, 2012
- Date of Event
- February 22, 2012
- Report Date
- May 30, 2012
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K961481
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
THE INVESTIGATION COULD NOT DETERMINE A SPECIFIC ROOT CAUSE. A GENERAL REAGENT OR INSTRUMENT ISSUE IS NOT OBVIOUS AS ALL CALIBRATION AND QUALITY CONTROL DATA WAS ACCEPTABLE. A SYSTEM ISSUE WAS NOT SUSPECTED AS ALL PERFORMANCE TESTING WAS SUCCESSFUL. THE PATIENT WAS NOT ADVERSELY AFFECTED.
THE USER RECEIVED A QUESTIONABLE INTACT HUMAN CHORIONIC GONADOTROPIN + THE SS-SUBUNIT (HCG+SS) RESULT FOR ONE PATIENT SAMPLE. THE INITIAL RESULT FROM A "POUR OFF" TUBE WAS >10000 MIU/ML WITH A DATA FLAG ON ANALYTICAL E MODULE SERIAL NUMBER (B)(4). THE SAMPLE WAS AUTOMATICALLY REPEATED ON THIS E MODULE WITH A 1:100 DILUTION AND THE RESULT WAS 10 MIU/ML WITH A DATA FLAG. THIS RESULT WAS REPORTED OUTSIDE THE LABORATORY. THE USER QUESTIONED THE RESULT AND RERAN THE ORIGINAL SAMPLE ON ANALYTICAL E MODULE SERIAL NUMBER (B)(4) AND THE RESULT WAS >10,000 MIU/ML WITH A DATA FLAG. THE SAMPLE AUTOMATICALLY REPEATED WITH A 1:100 DILUTION AND THE RESULT WAS 113085 MIU/ML WITH A DATA FLAG. THIS WAS THE RESULT THE USER REPORTED AS CORRECT. THE ORIGINAL "POUR OFF" TUBE WAS REPEATED AND THE RESULT WAS >10000 MIU/ML WITH A DATA FLAG. THE SAMPLE WAS AUTOMATICALLY REPEATED WITH A 1:100 DILUTION AND THE RESULT WAS 116712 MIU/ML. THE PATIENT WAS NOT ADVERSELY AFFECTED. THE HCG+SS REAGENT LOT NUMBER WAS 16250103 WITH AN EXPIRATION DATE OF 07/31/2012. THE FIELD SERVICE REPRESENTATIVE COULD NOT DETERMINE A CAUSE. HE ADJUSTED THE LIQUID LEVEL DETECTION (LLD) VOLTAGE, RAN PERFORMANCE TESTING AND TESTED SAMPLES AND ALL "POUR OFFS." HE OBSERVED THE SYSTEM WAS PERFORMING CORRECTLY AND AS PER SPECIFICATION. ALL SYSTEM CHECKS WERE SUCCESSFUL. THE USER RAN QUALITY CONTROL WITH ALL RESULTS ACCEPTABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANALYTICAL E MODULE | IMMUNOCHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |