FDA Adverse Event Summary report: N

INGENUITY TF PET/MR

MDR report key: 2483586 · Received February 28, 2012

Report

Report Number
1525965-2012-00004
Date Received
February 28, 2012
Date of Event
January 24, 2012
Report Date
February 3, 2012
Manufacturer
PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
Product Code
OUO
PMA / PMN Number
K103483
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THERE IS A REMOTE POTENTIAL FOR FALSE-NEGATIVE INTERPRETATION IN PTS WITH NEOPLASM THAT COULD RESULT IN LACK OF TREATMENT LEADING TO THE NECESSITY FOR MEDICAL INTERVENTION TO PREVENT PERMANENT IMPAIRMENT. THERE WAS NO REPORT OF HARM TO A PT IN THIS CASE AND THE HARM IS NOT EXPECTED TO OCCUR DUE TO TRAINED INTERPRETING PHYSICIAN RECOGNIZING IMAGE ARTIFACT, AND DEFAULT DISPLAY OF NON-AC AND ATTENUATION MAP FOR COMPARISON. INSTRUCTIONS FOR USE INCLUDES REVIEW OF NON-ATTENUATION CORRECTION IMAGES (459800153451 REV. B P18). THIS REPORTED EVENT IS CURRENTLY UNDER INVESTIGATION. A FOLLOW-UP REPORT WILL BE SENT WHEN THE INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

PHILIPS RECEIVED A CUSTOMER COMPLAINT STATING A PHOTOPENIC ARTIFACT WAS DETECTED THROUGH THE LOWER LUNGS ON THE MAGNETIC RESONANCE ATTENUATION CORRECTED (MRAC) IMAGE. THE CUSTOMER ALSO NOTICED LOW UPTAKE IN THE INFERIOR WALL OF THE HEART THAT WAS ALSO DETECTED ON THE NON-ATTENUATION CORRECTIVE (NAC) IMAGES. THE ARTIFACT IS NOT PRESENT ON THE MR MAP OR NAC DATASET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INGENUITY TF PET/MR INGENUITY TF PET/MR OUO PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC. 882380

Patients

Seq Age Sex Outcome Treatment
1