DUROM US ACETABULAR COMPONENT 56/50 P
Report
- Report Number
- 9613350-2012-00279
- Event Type
- Injury
- Date Received
- March 6, 2012
- Date of Event
- April 2, 2009
- Report Date
- February 15, 2012
- Manufacturer
- ZIMMER INC.
- Product Code
- KWB
- Removal / Correction Number
- 9613350-07/15/2008-001C
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE MFR DID NOT RECEIVE DEVICES, X-RAYS, OR OTHER SOURCE DOCUMENTS FOR REVIEW. WHERE LOT NUMBERS WERE RECEIVED FOR THE DEVICES, THE DEVICE HISTORY RECORDS WERE REVIEWED AND FOUND TO BE CONFORMING. WHILE A CAUSE FOR THIS SPECIFIC CLINICAL EVENT CANNOT BE ASCERTAINED FROM THE INFO PROVIDED, THE COMMON CLINICAL PRESENTATION AND THE DATE OF THE ORIGINAL IMPLANTATION SUGGEST THAT THIS CASE IS RELATED TO THE ISSUES FOR WHICH ZIMMER IMPLEMENTED A CORRECTION IN JULY 2008. SHOULD ADD'L INFO BECOME AVAILABLE AND/OR THE DEVICE(S) BE RETURNED FOR EVAL AND AN INVESTIGATION RESULT BE AVAILABLE, THAT CHANGES THIS ASSESSMENT, AN AMENDED MEDICAL DEVICE REPORT WILL BE FILED. THE NEED FOR FURTHER CORRECTIVE MEASURES IS NOT INDICATED AT THIS TIME AND ZIMMER (B)(4) CONSIDERS THIS CASE AS CLOSED. (B)(4).
IT WAS REPORTED BY PT'S COUNSEL THAT THE PT RECEIVED AN IMPLANT ON (B)(6) 2008 AND WAS REVISED ON (B)(6) 2009 DUE TO PAIN, MULTIPLE DISLOCATIONS, AND LOOSENING. THE SURGEON FOUND NO EVIDENCE OF BONY INGROWTH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DUROM US ACETABULAR COMPONENT 56/50 P | DUROM ACETABULAR COMPONENT AND METASUL | KWB | ZIMMER INC. | 2432801 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Hospitalization| R |