FDA Adverse Event Injury Summary report: N

DAVINCI XI

MDR report key: 24829031 · Received April 9, 2026

Report

Report Number
2955842-2026-20787
Event Type
Injury
Date Received
April 9, 2026
Date of Event
March 14, 2026
Report Date
April 9, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110720
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED EVENT WAS ADDRESSED WITH PHONE SUPPORT. THE INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) RECOMMENDED RESEATING THE STERILE ADAPTER AND SWAPPING THE INSTRUMENT. LATER, THE CUSTOMER CONFIRMED THAT ISSUE WAS RESOLVED AND STATED THAT THE HEAD SENSOR HAD NOT BEEN BLOCKED PROPERLY. NO SITE VISIT WAS CONDUCTED. THE SYSTEM WAS WORKING PROPERLY, AND NO ADDITIONAL ACTION WAS REQUIRED AS THE ISSUE WAS RESOLVED. NO PRODUCT IS EXPECTED TO BE RETURNED FOR EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING AN UNSPECIFIED DA VINCI-ASSISTED PROCEDURE, THE SURGEON WAS UNABLE TO TAKE CONTROL OF THE ROBOTIC INSTRUMENTS AND NOTED A PATIENT INJURY. AN INTUITIVE SURGICAL INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) WAS CONSULTED AND RECOMMENDED RESEATING THE STERILE ADAPTER ON THE DRAPE AND SWAPPING THE INSTRUMENT; HOWEVER, THESE ACTIONS DID NOT RESOLVE THE ISSUE. FURTHER INVESTIGATION DETERMINED THAT THE LOSS OF INSTRUMENT CONTROL WAS DUE TO THE SURGEON¿S HEAD NOT FULLY BLOCKING THE HIGH-RESOLUTION STEREO VIEWER (HRSV), WHICH PREVENTED THE SYSTEM FROM RECOGNIZING THE SURGEON'S PRESENCE AND ENABLING INSTRUMENT MANIPULATION. THE FOLLOWING INFORMATION IS UNKNOWN: THE CAUSE OF THE PATIENT INJURY, IF ANY BLEEDING OCCURRED, AND THE MEDICAL INTERVENTION PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED; HOWEVER, NO RESPONSE HAS BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
183539 DAVINCI XI PATIENT SIDE CART, 4-ARM NAY INTUITIVE SURGICAL, INC 380652-58 N/A 00886874110720

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other DA VINCI INSTRUMENTS AND ACCESSORIES