FDA Adverse Event Malfunction Summary report: N

DIGITAL RADIOGRAM CXDI-55G

MDR report key: 2482721 · Received February 27, 2012

Report

Report Number
1000181430-2012-00002
Event Type
Malfunction
Date Received
February 27, 2012
Date of Event
April 26, 2010
Report Date
February 27, 2012
Manufacturer
CANON, INC.
Product Code
MQB
PMA / PMN Number
K091435
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE FOLLOWING INFO NEED TO BE MADE AS COUNTERMEASURES. SCREW TYPE (FOR THE TWO LOCATIONS) NEED TO BE CHANGED FROM L = 4MM TO 6MM + WASHER + SPRING WASHER. SCREW TIGHTENING TORQUE NEED TO BE INCREASED FROM 3KGF CM TO 4KGF CM. THREAD LOCKING NEED TO BE APPLIED TO THE SCREW HEAD AND PCB.

Description of Event or Problem · 1

SCREWS HOLDING THE 55REF CIRCUIT BOARD ARE LOOSENING AND ROAMING INSIDE OF THE SEALED SENSOR. THIS CAN ALSO CAUSE THE CIRCUIT BOARD TO SHORT CIRCUIT, CAUSE IMAGE FAILURE, AND ERROR INTERMITTENTLY. IN SOME CASES, IT CAN CAUSE THE BOARD TO PERMANENTLY FAIL. WITH IMAGE FAILURE, IT COULD CAUSE EXCESSIVE RADIATION EXPOSURE TO THE PT WITH NO RESULTING IMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIGITAL RADIOGRAM CXDI-55G SOLID STATE X-RAY IMAGER MQB CANON, INC. CXDI-55G NA

Patients

Seq Age Sex Outcome Treatment
1 Other