FDA Adverse Event
Malfunction
Summary report: N
DIGITAL RADIOGRAM CXDI-55G
MDR report key: 2482721
·
Received February 27, 2012
Report
- Report Number
- 1000181430-2012-00002
- Event Type
- Malfunction
- Date Received
- February 27, 2012
- Date of Event
- April 26, 2010
- Report Date
- February 27, 2012
- Manufacturer
- CANON, INC.
- Product Code
- MQB
- PMA / PMN Number
- K091435
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE FOLLOWING INFO NEED TO BE MADE AS COUNTERMEASURES. SCREW TYPE (FOR THE TWO LOCATIONS) NEED TO BE CHANGED FROM L = 4MM TO 6MM + WASHER + SPRING WASHER. SCREW TIGHTENING TORQUE NEED TO BE INCREASED FROM 3KGF CM TO 4KGF CM. THREAD LOCKING NEED TO BE APPLIED TO THE SCREW HEAD AND PCB.
Description of Event or Problem · 1
SCREWS HOLDING THE 55REF CIRCUIT BOARD ARE LOOSENING AND ROAMING INSIDE OF THE SEALED SENSOR. THIS CAN ALSO CAUSE THE CIRCUIT BOARD TO SHORT CIRCUIT, CAUSE IMAGE FAILURE, AND ERROR INTERMITTENTLY. IN SOME CASES, IT CAN CAUSE THE BOARD TO PERMANENTLY FAIL. WITH IMAGE FAILURE, IT COULD CAUSE EXCESSIVE RADIATION EXPOSURE TO THE PT WITH NO RESULTING IMAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIGITAL RADIOGRAM CXDI-55G | SOLID STATE X-RAY IMAGER | MQB | CANON, INC. | CXDI-55G | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |