FDA Adverse Event Malfunction Summary report: N

NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE

MDR report key: 2482298 · Received March 2, 2012

Report

Report Number
1811755-2012-00732
Event Type
Malfunction
Date Received
March 2, 2012
Date of Event
November 9, 2011
Report Date
December 15, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSPECTION DISCOVERED A CUT IN THE POWER CORD, WHICH EXPOSED BARE WIRES. THE CAUSE OF THE CORD DAMAGE IS UNK, BUT MAY BE THE RESULT OF MISHANDLING. THE MFR SERVICE TECH INSTALLED A NEW POWER PLUG AT THE POINT WHERE THE POWER CORD AS CUT. THE ROVER WAS FUNCTIONALLY TESTED AND RETURNED TO USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN EVAL CONDUCTED BY A MFR FIELD SERVICE TECH, EXPOSED BARE WIRES WERE OBSERVED FROM A CUT IN THE ROVER POWER CORD. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE FYD STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK