ARCHITECT STAT HIGH SENSITIVE TROPONIN-I
Report
- Report Number
- 3005094123-2026-00168
- Event Type
- Malfunction
- Date Received
- April 8, 2026
- Date of Event
- March 30, 2026
- Report Date
- April 8, 2026
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- MMI
- UDI-DI
- 00380740165093
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 03P25, THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 02R98, TROPONIN-I STAT HIGH SENSITIVITY, WITH 510K/PMA/BLA NUMBER K191595. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT STAT HIGH SENSITIVE TROPONIN-I FOR A 51 YEAR OLD MALE. THE SAMPLE WAS REPEATED USING PEG PRECIPITATION WITH LOWER RESULTS. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE =34.2 PG/ML): SID (B)(6), INITIAL TROPONIN RESULT= 426.99 PG/ML; REPEAT RESULT= 414.01 PG/ML; REPEAT RESULT USING PEG PRECIPITATION= 1.44 PG/ML THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 881160 | ARCHITECT STAT HIGH SENSITIVE TROPONIN-I | IMMUNOASSAY METHOD, TROPONIN SUBUNIT | MMI | ABBOTT IRELAND DIAGNOSTICS DIVISION | 82349UD01 | 00380740165093 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Male | ARC I2000SR INST, 03M74-02, (B)(6. |