FDA Adverse Event Malfunction Summary report: N

ARCHITECT STAT HIGH SENSITIVE TROPONIN-I

MDR report key: 24815979 · Received April 8, 2026

Report

Report Number
3005094123-2026-00168
Event Type
Malfunction
Date Received
April 8, 2026
Date of Event
March 30, 2026
Report Date
April 8, 2026
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
MMI
UDI-DI
00380740165093
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 03P25, THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 02R98, TROPONIN-I STAT HIGH SENSITIVITY, WITH 510K/PMA/BLA NUMBER K191595. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT STAT HIGH SENSITIVE TROPONIN-I FOR A 51 YEAR OLD MALE. THE SAMPLE WAS REPEATED USING PEG PRECIPITATION WITH LOWER RESULTS. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE =34.2 PG/ML): SID (B)(6), INITIAL TROPONIN RESULT= 426.99 PG/ML; REPEAT RESULT= 414.01 PG/ML; REPEAT RESULT USING PEG PRECIPITATION= 1.44 PG/ML THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
881160 ARCHITECT STAT HIGH SENSITIVE TROPONIN-I IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ABBOTT IRELAND DIAGNOSTICS DIVISION 82349UD01 00380740165093

Patients

Seq Age Sex Outcome Treatment
1 51 YR Male ARC I2000SR INST, 03M74-02, (B)(6.