ADVIA CENTAUR XP HCV ASSAY
Report
- Report Number
- 1219913-2012-00089
- Event Type
- Malfunction
- Date Received
- March 7, 2012
- Date of Event
- February 1, 2012
- Report Date
- February 10, 2012
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC.
- Product Code
- MZO
- PMA / PMN Number
- P030056
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE CAUSE OF THE INCREASE POSITIVE RESULTS FOR LOT # 228 IS BEING INVESTIGATED. THE IFU STATES IN THE INTERPRETATION OF RESULTS SECTION: "SAMPLES WITH A CALCULATED VALUE GREATER THAN OR EQUAL TO 1.00 INDEX VALUE ARE CONSIDERED REACTIVE FOR IGG ANTIBODIES TO HCV. IT IS RECOMMENDED THAT THE SAMPLE BE REPEATED IN DUPLICATE. IF 2 OF THE 3 SAMPLE RESULTS ARE LESS THAN 0.80 INDEX VALUE, THE SAMPLE IS CONSIDERED NONREACTIVE. IF 2 OF THE 3 SAMPLE RESULTS ARE GREATER THAN OR EQUAL TO 1.00 INDEX VALUE, THE SAMPLE IS CONSIDERED REACTIVE AND SUPPLEMENTAL TESTING OF THE SAMPLE IS RECOMMENDED. IF 2 OF THE 3 SAMPLE RESULTS ARE GREATER THAN OR EQUAL TO 0.80 INDEX VALUE AND LESS THAN 1.00 INDEX VALUE, SUPPLEMENTAL TESTING OF THE SAMPLE IS RECOMMENDED."
(B)(4). INTERNAL STUDIES HAVE CONFIRMED THAT THIS INCREASED REACTIVE RATE IS NOT WITHIN THE RESOLVED SPECIFICITY AS STATED IN THE PERFORMANCE CHARACTERISTICS SECTION OF THE INSTRUCTIONS FOR USE, DISTRIBUTED OUTSIDE THE US: (B)(4). AS A RESULT, URGENT FIELD SAFETY NOTICE, (B)(4), WAS ISSUED TO SIEMENS' CUSTOMERS OUTSIDE THE US APRIL, 2012.
FALSE POSITIVE RESULTS FROM WITHIN AN EQUIVALENT POPULATION WAS OBSERVED ON THE ADVIA CENTAUR XP HCV LOT # 228 AT THIS FACILITY. THE POSITIVE RESULTS WERE REPORTED BY THE CUSTOMER AND CONFIRMATION TESTING WAS NOT PERFORMED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ADVIA CENTAUR XP HCV RESULTS.