AUTOVENT 4000CPAP W/ALARM
Report
- Report Number
- 1924066-2012-00003
- Event Type
- Malfunction
- Date Received
- February 28, 2012
- Date of Event
- February 13, 2012
- Report Date
- February 23, 2012
- Manufacturer
- ALLIED HEALTHCARE PRODUCTS, INC.
- Product Code
- BTL
- PMA / PMN Number
- K070636
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
UNIT WAS RETURNED; (BPM) BREATHES PER MINUTE WERE TESTED ON UNIT. ALLIED COULD NOT DUPLICATE COMPLAINT; RAN UNIT FOR 3 TO 4 HOURS; NO CHANGE IN BPM. IT APPEARS THE PATIENT MAY HAVE STARTED TO BREATHE SPONTANEOUSLY. IF A PATIENT BEGINS TO BREATHE SPONTANEOUSLY, THE BREATH ASSIST OF THE VENTILATOR WILL ADJUST THE BPM OF THE PATIENT'S BREATH RATE. THE AUTOVENT 4000 OPERATION MANUAL ASSISTS THE USER WHEN THIS SITUATION OCCURS: OPERATING THE AUTOVENT: (VENTILATION). SPONTANEOUS BREATHING BY THE PATIENT: SHOULD THE PATIENT BEGIN TO BREATHE SPONTANEOUSLY THE AUTOVENT 4000CPAP WILL SENSE THIS BREATH AND DELIVER THE SET TIDAL VOLUME AT THE SET INSPIRATORY RATE. THE BREATH TIMING WILL BE RESET BASED ON THE SELECTED BPM RATE. FOR EXAMPLE, IF 10 BPM WAS SELECTED THE NEXT BREATH WILL BE DELIVERED 6 SECONDS AFTER THE START OF THE SPONTANEOUS BREATH.
IT WAS REPORTED THE PARAMEDIC HAD A PATIENT EN ROUTE ON AN HOUR LONG TRANSPORT USING AN AUTOVENT. THE UNIT WAS SET FOR 12 BPM BUT STARTED GIVING BREATHES AT ABOUT 24 BPM. IT SEEMED TO THE PARAMEDIC THE UNIT WAS JUST BLOWING NON-STOP. HE UNPLUGGED THE UNIT TRYING TO RESET IT; PLUGGED IT BACK IN; UNIT WORKED FOR ABOUT 30 SECONDS. AS A RESULT THE PARAMEDIC HAD TO MANUALLY BAG THE PATIENT FOR MOST OF THE TRANSPORT. THIS MADE IT DIFFICULT TO CONTROL DOPHAMEAN AND RESUSCITATE AT THE SAME TIME. AS A RESULT THE PATIENT'S BLOOD PRESSURE SHIFTED FROM HIGH TO LOW. NO INFORMATION WAS GIVEN ON END RESULT OF PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOVENT 4000CPAP W/ALARM | VENTILATOR | BTL | ALLIED HEALTHCARE PRODUCTS, INC. | L763CPAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |