FDA Adverse Event Malfunction Summary report: N

VISIAN ICL (IMPLANTABLE COLLAMER LENS)

MDR report key: 2480284 · Received February 28, 2012

Report

Report Number
2023826-2012-00153
Event Type
Malfunction
Date Received
February 28, 2012
Date of Event
January 31, 2012
Report Date
January 31, 2012
Manufacturer
STAAR SURGICAL CO.
Product Code
MTA
PMA / PMN Number
P030016
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (METHOD) - LENS WORK ORDER SEARCH. (RESULTS) - VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND ALMOST WHOLE PART OF ONE HAPTIC AND PIECE OF OPTIC IS TORN OFF AND MISSING. LENS WAS RETURNED IN LIQUID. A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE SAME WORK ORDER. (CONCLUSIONS) - BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4).

Description of Event or Problem · 1

THE REPORTER STATED THE SURGEON USED AN MICL 13.2MM IMPLANTABLE COLLAMER LENS. THE SURGEON WAS ADVANCING THE LENS INTO THE RIGHT EYE INCISION. IT WAS PARTIALLY IN THE EYE WHEN THE LENS GOT CAUGHT AND TORE AS IT WAS COMING OUT OF THE INJECTOR. THE SURGEON WAS ABLE TO PULL THE LENS OUT OF THE EYE WITH THE LENS STILL PARTIALLY INSIDE THE INJECTOR/CARTRIDGE. THERE WAS NO PATIENT INJURY. ANOTHER SAME MODEL AND SIZE LENS WAS IMPLANTED. NO LOT NUMBERS WERE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VISIAN ICL (IMPLANTABLE COLLAMER LENS) INTRAOCULAR LENS MTA STAAR SURGICAL CO. MICL13.2 NA

Patients

Seq Age Sex Outcome Treatment
1 46 YR CARTRIDGE: MODEL SFC-45 FP| INJECTOR: MODEL MSI-PF| FOAM TIP PLUNGER