FDA Adverse Event Malfunction Summary report: N

DUAL ARM ARTICULATING EQUIPMENT BOOM

MDR report key: 2480203 · Received February 28, 2012

Report

Report Number
2031963-2012-00036
Event Type
Malfunction
Date Received
February 28, 2012
Date of Event
January 31, 2012
Report Date
January 31, 2012
Manufacturer
STRYKER COMMUNICATIONS
Product Code
BRY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO REPORTED PATIENT INVOLVEMENT, THEREFORE, NO PATIENT DATA EXISTS. SERIAL NUMBER IS UNKNOWN. ADDITIONAL ATTEMPTS WILL BE MADE FOR THIS INFORMATION. THE DEVICE HAS NOT BEEN EVALUATED AT THE TIME OF THIS REPORT. EVALUATION SUMMARY: THE STRYKER EQUIPMENT DELIVERY SYSTEM (EDS) IS INTENDED FOR USE AS A CEILING-MOUNTED DEVICE FOR SUPPORTING OR POSITIONING EQUIPMENT IN THE SURGICAL SUITE AND/OR DELIVERS GASES OR ELECTRICITY TO THESE DEVICES. THE MMP-200 ARM-SET IS AN ARTICULATING ARM-SET THAT FEATURES A MOTOR UNDER THE HORIZONTAL ARM. THE MMP-200 OFFERS VERTICAL ARTICULATION IN CONJUNCTION WITH THE HORIZONTAL ARTICULATIONS POSSIBLE WITH THE OSC-400 AND OSC-600 ARM-SETS. THE CUSTOMER HAS NOT SCHEDULED AN ON-SITE EVALUATION OR REPAIR AT THE TIME OF THIS REPORT, THUS ROOT CAUSES OF THE POTENTIAL MALFUNCTIONS ARE UNKNOWN. HOWEVER, BASED ON THE INFORMATION KNOWN AT THIS POINT IN TIME, THERE IS NOT ENOUGH INFORMATION TO RULE OUT THE POTENTIAL HEALTH RISK OF THE BOOM DROPPING ENOUGH TO CAUSE THE EQUIPMENT TO FALL OFF THE SHELF. HOWEVER, IN THIS INSTANCE THERE WERE NO PATIENTS INVOLVED AND NO ADVERSE EVENTS WERE REPORTED. THIS NON-CONFORMANCE WILL BE MONITORED. THIS IS NOT A SINGLE USE DEVICE.

Description of Event or Problem · 1

(B)(4). IT WAS REPORTED THAT WHEN TRYING TO ELEVATE THE LOWER ARM OF THE EQUIPMENT BOOM, IT DROPPED ABOUT 10 INCHES AND THEN CAUGHT ITSELF. THERE WAS NO REPORTED PATIENT INVOLVEMENT AND NO REPORTED ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DUAL ARM ARTICULATING EQUIPMENT BOOM BRY STRYKER COMMUNICATIONS

Patients

Seq Age Sex Outcome Treatment
1