FDA Adverse Event Malfunction Summary report: N

DIGITALDIAGNOST DUAL DETECTOR

MDR report key: 2480072 · Received February 28, 2012

Report

Report Number
3003768251-2012-00005
Event Type
Malfunction
Date Received
February 28, 2012
Report Date
February 3, 2012
Manufacturer
PHILIPS MEDICAL SYSTEMS DMC GMBH
Product Code
MQB
PMA / PMN Number
K982796
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER INVESTIGATION IDENTIFIED THAT THE PHYSICIAN WAS NOT USING THE SYSTEM CORRECTLY; THE AUTOMATIC STITCHING PROCESS CAN LEAD TO AN INCORRECT IMAGE DUE TO PATIENT MOVEMENTS, PATIENT BREATHING AND RULER MOVEMENT. AS PART OF THE AUTOMATED STITCHING WORKFLOW IT IS IMPORTANT THAT THE PHYSICIAN REVIEWS THE INDIVIDUAL IMAGES AGAINST THE COMPOSITE IMAGE. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED A STITCHING PROBLEM, WHICH CAN POTENTIALLY LEAD TO A MISDIAGNOSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIGITALDIAGNOST DUAL DETECTOR MQB PHILIPS MEDICAL SYSTEMS DMC GMBH 712022

Patients

Seq Age Sex Outcome Treatment
1