FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARIN 56 UNITS

MDR report key: 24794084 · Received April 6, 2026

Report

Report Number
1917413-2026-00225
Event Type
Malfunction
Date Received
April 6, 2026
Date of Event
March 10, 2026
Report Date
March 11, 2026
Manufacturer
BECTON DICKINSON & CO (FRANKLIN LAKES)
Product Code
JKA
UDI-DI
30382903679608
PMA / PMN Number
K945952
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K954592. INVESTIGATION SUMMARY: BD RECEIVED 2 PHOTOS FOR INVESTIGATION, AND THESE SHOW AIR BUBBLES INSIDE THE GEL AND FOREIGN MATTER. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODES: FOREIGN MATTER - NON-BIOLOGICAL AND GEL AIRBUBBLES. NO DEFINITIVE ROOT CAUSE COULD BE ESTABLISHED FOR THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USING BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARIN 56 UNITS, FOREIGN MATTER WAS FOUND IN 300 TUBES AND AIR BUBBLES WERE IN 1600. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
857182 BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARIN 56 UNITS TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION JKA BECTON DICKINSON & CO (FRANKLIN LAKES) 5225554 30382903679608

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown