BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARIN 56 UNITS
Report
- Report Number
- 1917413-2026-00225
- Event Type
- Malfunction
- Date Received
- April 6, 2026
- Date of Event
- March 10, 2026
- Report Date
- March 11, 2026
- Manufacturer
- BECTON DICKINSON & CO (FRANKLIN LAKES)
- Product Code
- JKA
- UDI-DI
- 30382903679608
- PMA / PMN Number
- K945952
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K954592. INVESTIGATION SUMMARY: BD RECEIVED 2 PHOTOS FOR INVESTIGATION, AND THESE SHOW AIR BUBBLES INSIDE THE GEL AND FOREIGN MATTER. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODES: FOREIGN MATTER - NON-BIOLOGICAL AND GEL AIRBUBBLES. NO DEFINITIVE ROOT CAUSE COULD BE ESTABLISHED FOR THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED PRIOR TO USING BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARIN 56 UNITS, FOREIGN MATTER WAS FOUND IN 300 TUBES AND AIR BUBBLES WERE IN 1600. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 857182 | BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARIN 56 UNITS | TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION | JKA | BECTON DICKINSON & CO (FRANKLIN LAKES) | 5225554 | 30382903679608 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |