FDA Adverse Event Injury Summary report: N

UNK GEL BREAST IMPLANT

MDR report key: 24791052 · Received April 6, 2026

Report

Report Number
9617229-2026-06134
Event Type
Injury
Date Received
April 6, 2026
Date of Event
March 11, 2026
Report Date
April 6, 2026
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT RUPTURE WAS RECEIVED ON (B)(6) 2026, WITH LOT NUMBER 1170877. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION. PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS REQUIRED: - RUPTURE: OBSERVED BROKEN DEVICE THROUGH MICROSCOPIC INSPECTION ASSESSED AS SURGICAL DAMAGE AND OBSERVED A MISSING PIECE OF SHELL THROUGH MICROSCOPIC INSPECTION ASSESSED AS INCONCLUSIVE. NO FURTHER ACTIONS ARE REQUIRED AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS (NON-PENETRATING NICK) ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS. ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: D.9., H.2., H.3., H.6.

Additional Manufacturer Narrative · 0

FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE LEFT SIDE. THE DEVICE WAS EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE LEFT SIDE. THE DEVICE WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
849321 UNK GEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) NI

Patients

Seq Age Sex Outcome Treatment
1 47 YR Female Required Intervention