FDA Adverse Event Malfunction Summary report: N

CD HORIZON® SOLERA® VOYAGER¿ SPINAL SYSTEM

MDR report key: 24787374 · Received April 6, 2026

Report

Report Number
3003120897-2026-00647
Event Type
Malfunction
Date Received
April 6, 2026
Date of Event
December 29, 2025
Report Date
April 6, 2026
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC.
Product Code
HWE
UDI-DI
00643169961623
PMA / PMN Number
K170679
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS PART # 6550017, LOT # K25B1032 VISUAL INSPECTION CONFIRMED THE END OF THE TAB EXTENDER THAT INTERFACES WITH THE BREAK OFF SCREW HAS BEEN BENT. THE DAMAGE TO THE EXTENDER IS CONSISTENT WITH OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY FOR TH12 DISH. IT WAS REPORTED THAT DURING THE FINAL TAB BREAKING AFTER V5 WAS FULLY TIGHTENED, FORCE WAS APPLIED TO BREAK THE TAB BEFORE THE TAB BREAKER WAS FULLY INSERTED, RESULTING IN THE TAB BECOMING BENT. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
851975 CD HORIZON® SOLERA® VOYAGER¿ SPINAL SYSTEM INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED M HWE MEDTRONIC SOFAMOR DANEK USA, INC. 6550017 K25B1032 00643169961623

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown