FDA Adverse Event Malfunction Summary report: N

KEEPSAFE DELUXE NO POWER

MDR report key: 2478477 · Received February 16, 2012

Report

Report Number
2020362-2012-00077
Event Type
Malfunction
Date Received
February 16, 2012
Report Date
November 13, 2012
Manufacturer
J. T. POSEY CO.
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) BATTERY LEAKAGE, CORROSION. EVALUATION OF THE RETURNED PRODUCT FOUND THE UNIT DID NOT POWER UP WHEN TESTED WITH BATTERIES. FURTHER INVESTIGATION REVEALED THE BATTERY DOOR HAS CORROSION AND BATTERY LEAKAGE. THE UNIT POWERED UP WITH A POWER SUPPLY AND APPEARS TO WORK. NOTE: POSEY INSTRUCTIONS FOR USE HAS A WARNING STATEMENT: TAKE CARE WHEN INSTALLING NEW BATTERIES. THE ALARM WILL NOT WORK IF BATTERIES ARE INSTALLED IMPROPERLY. DO NOT MIX OLD AND NEW BATTERIES, OR BATTERY BRANDS WITHIN A BATTERY PACK. ALWAYS INSTALL A COMPLETELY NEW SET OF BATTERIES. DO NOT ALLOW BATTERIES TO DEPLETE WHILE IN THE ALARM. DEPLETED BATTERIES MAY LEAK AND CORRODE, CAUSING DAMAGE TO THE ELECTRONICS AND RELIABILITY. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED THE ALARM HAS BATTERY CORROSION. THERE WAS NO PATIENT INCIDENT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KEEPSAFE DELUXE NO POWER KMI J. T. POSEY CO. 8374NP NA

Patients

Seq Age Sex Outcome Treatment
1 NA