FDA Adverse Event Malfunction Summary report: N

SITTER II

MDR report key: 2478468 · Received February 16, 2012

Report

Report Number
2020362-2012-00076
Event Type
Malfunction
Date Received
February 16, 2012
Report Date
January 16, 2012
Manufacturer
J. T. POSEY CO.
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: EVALUATION OF THE RETURNED PRODUCT FOUND THE ALARM POWERS UP BUT THE LOW BATTERY TEST AND THE VOICE MESSAGE FUNCTION FAILED. THE DISPLAY SCREEN HAS A SUBSTANCE ON THE INSIDE OF THE SCREEN AND THE NUMBERS ARE NOT VISIBLE. SENSOR FUNCTION ONLY WORKS ON INITIAL INSTALLATION OF THE BATTERIES AND NOT AFTER THE BUTTONS ARE PRESSED. THE SELECT UP AND DOWN BUTTONS DO NOT WORK. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE ALARM HAS POWER BUT THE DISPLAY IS BLANK. ALL OTHER FUNCTIONS OF THE ALARM WORK FINE. THERE WAS NO PT INCIDENT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SITTER II KMI J. T. POSEY CO. 8281 NA

Patients

Seq Age Sex Outcome Treatment
1 NI LOT NUMBER: UNK| OVER-THE-MATTRESS SENSOR PAD MODEL: 8307,