FDA Adverse Event
Malfunction
Summary report: N
SITTER II
MDR report key: 2478468
·
Received February 16, 2012
Report
- Report Number
- 2020362-2012-00076
- Event Type
- Malfunction
- Date Received
- February 16, 2012
- Report Date
- January 16, 2012
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). RESULTS: EVALUATION OF THE RETURNED PRODUCT FOUND THE ALARM POWERS UP BUT THE LOW BATTERY TEST AND THE VOICE MESSAGE FUNCTION FAILED. THE DISPLAY SCREEN HAS A SUBSTANCE ON THE INSIDE OF THE SCREEN AND THE NUMBERS ARE NOT VISIBLE. SENSOR FUNCTION ONLY WORKS ON INITIAL INSTALLATION OF THE BATTERIES AND NOT AFTER THE BUTTONS ARE PRESSED. THE SELECT UP AND DOWN BUTTONS DO NOT WORK. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE ALARM HAS POWER BUT THE DISPLAY IS BLANK. ALL OTHER FUNCTIONS OF THE ALARM WORK FINE. THERE WAS NO PT INCIDENT OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SITTER II | KMI | J. T. POSEY CO. | 8281 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | LOT NUMBER: UNK| OVER-THE-MATTRESS SENSOR PAD MODEL: 8307, |