FORESIGHT IQ
Report
- Report Number
- 2015691-2026-13832
- Event Type
- Malfunction
- Date Received
- April 3, 2026
- Date of Event
- March 16, 2026
- Report Date
- April 27, 2026
- Manufacturer
- EDWARDS LIFESCIENCES IRVINE
- Product Code
- DQE
- UDI-DI
- 00690103219616
- PMA / PMN Number
- K242451
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
PER REVIEW OF THE IMAGES PROVIDED BY CUSTOMER, IT'S CONFIRMED IMAGE 1 SHOWS A FORESIGHT THB OF 9.2G/DL AND IMAGE 2 SHOWS ARTERIAL BLOOD LAB RESULTS WITH A HB OF 6.8G/DL. THE DEVICE WAS NOT RETURNED FOR EVALUATION. WITHOUT THE RETURN OF THE UNIT, IT IS NOT POSSIBLE TO DETERMINE IF SOME DAMAGE OR DEFECT EXISTED ON THE UNIT THAT COULD HAVE CONTRIBUTED TO THE EVENT. IT IS NOT KNOWN IF SOME PROCEDURAL FACTORS MAY HAVE CONTRIBUTED TO THE EVENT. A DEVICE HISTORY RECORD REVIEW WAS UNABLE TO BE COMPLETED AS THE LOT NUMBER WAS NOT PROVIDED.
ADDITIONAL FDA PRODUCT CODES INCLUDE: DSB- PLETHYSMOGRAPH, IMPEDANCE. DXN- SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE. FLL- CONTINUOUS MEASUREMENT THERMOMETER. MUD- OXIMETER, TISSUE SATURATION. QAQ- ADJUNCTIVE PREDICTIVE CARDIOVASCULAR INDICATOR. QEM- CEREBRAL OXIMETER. QMS- ADJUNCTIVE OPEN LOOP FLUID THERAPY RECOMMENDER. QNL- MEDIUM-TERM ADJUNCTIVE PREDICTIVE CARDIOVASCULAR INDICATOR. THE DEVICE EVALUATION IS ANTICIPATED. A SUPPLEMENTAL REPORT WILL BE FORTHCOMING WHEN THE INVESTIGATION IS COMPLETED. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS A PART OF THE MONTHLY REVIEW.
IT WAS REPORTED THAT DURING USE OF THE FORESIGHT SENSOR THE TOTAL HEMOGLOBIN WAS OFF BY MORE THAN 1 G. THERE WAS NO ALLEGATION OF PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 62398 | FORESIGHT IQ | CATHETER, OXIMETER, FIBER-OPTIC | DQE | EDWARDS LIFESCIENCES IRVINE | FSESLIQ | NI | 00690103219616 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |