FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK

MDR report key: 24778437 · Received April 3, 2026

Report

Report Number
9614033-2026-00021
Event Type
Malfunction
Date Received
April 3, 2026
Date of Event
February 23, 2026
Report Date
April 24, 2026
Manufacturer
BECTON DICKINSON DE MEXICO
Product Code
FMF
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: THREE PHOTOS WERE RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. A CRACK WAS OBSERVED ON THE WALL OF THE BARREL; THEREFORE, THE REPORTED INCIDENT COULD BE VERIFIED. A DEVICE HISTORY RECORD REVIEW FOUND NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING PRODUCTION OF THIS BATCH. BASED ON THE QUALITY TEAM'S INVESTIGATION, THE ROOT CAUSE IS RELATED TO MANUFACTURING. A PROJECT HAS BEEN STARTED TO FURTHER INVESTIGATE THE PROCESS. THIS INCIDENT HAS BEEN ADDED TO OUR DATABASE OF REPORTED INCIDENTS.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD SYRINGE 10 ML LL BNS BARREL CRACKED. COMPLAINT VIA EMAIL: WE WERE NOTIFIED OF A COMPLAINT FROM A CUSTOMER STATING SYRINGES CRACKED. PLEASE SEE THE BELOW DETAILS REGARDING THE REPORTED ISSUE. IF IT IS INDICATED THAT A SAMPLE IS AVAILABLE, PLEASE PROVIDE SHIPPING INSTRUCTIONS WITHIN 30 DAYS OR THE SAMPLE WILL BE DISPOSED OF. COMPLAINT #: (B)(4). DEFECT DESCRIPTION: SYRINGES CRACKED. PART #: 153239, PRODUCT DESCRIPTION: SYRINGE 10 ML LL BNS, VENDOR PART #: 304657, LOT #: 5069256. DATE REPORTED: 2/26/2026. SAMPLE RECEIVED: NO - BIOHAZARDOUS RESPONSE NEEDED: YES - INCIDENT REPORTED TO THE FDA AS MDR-30 DAY. REFERENCE NO. 1423395-2026-00044 PLEASE REFERENCE THE ABOVE COMPLAINT NUMBER IN ALL FUTURE COMMUNICATIONS. IF YOU HAVE QUESTIONS OR NEED ADDITIONAL INFORMATION, PLEASE REACH OUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
289281 BD LUER-LOK PISTON SYRINGE FMF BECTON DICKINSON DE MEXICO 5069256

Patients

Seq Age Sex Outcome Treatment
1