FDA Adverse Event Injury Summary report: N

EDISON

MDR report key: 24778090 · Received April 3, 2026

Report

Report Number
3027664504-2026-00011
Event Type
Injury
Date Received
April 3, 2026
Date of Event
March 4, 2026
Report Date
April 3, 2026
Manufacturer
HISTOSONICS, INC.
Product Code
QGM
UDI-DI
00850006962082
PMA / PMN Number
K241902
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

NO DEVICE MALFUNCTIONS OR OTHER NOTEWORTHY EVENTS OCCURRED DURING THE CASE. IN RESPONSE TO REPORTS OF ACUTE KIDNEY INJURY ASSOCIATED WITH HISTOTRIPSY, HISTOSONICS HAS INCORPORATED THE FOLLOWING WARNING INTO ITS LABELING: "CONSIDER PATIENT-SPECIFIC RISK FACTORS FOR ACUTE KIDNEY INJURY (AKI) AND/OR RENAL FAILURE BASED ON CLINICAL HISTORY AND PROCEDURAL SCENARIOS THAT COULD INCREASE STRESS ON KIDNEY FUNCTION, INCLUDING BUT NOT LIMITED TO HYDRATION STATUS, PLANNED TREATMENT VOLUME AND EXPECTED USE OF IMAGING CONTRAST AGENTS. CONSISTENT WITH OTHER LIVER-DIRECTED THERAPIES, TREATMENTS INVOLVING LARGE TREATMENT VOLUMES OR MULTIPLE TREATMENT SESSIONS MAY BE ASSOCIATED WITH AN INCREASED RISK OF AKI. EVALUATE RENAL RISK AND MONITOR FOR SIGNS OF AKI BEFORE AND AFTER TREATMENT."

Description of Event or Problem · 0

ON (B)(6) 2026, A 41 YEAR OLD FEMALE PATIENT WITH A HISTORY OF COLORECTAL CANCER WITH HEPATIC METASTASES RECEIVED HISTOTRIPSY TREATMENT TO ONE HEPATIC LESION IN SEGMENT VII FOR A TOTAL PLANNED TREATMENT VOLUME (PTV) OF 33.5 CC. THE PATIENT ALSO RECEIVED INTRAVENOUS FLUID RESUSCITATION, INCLUDING APPROXIMATELY 1 L OF LACTATED RINGER'S SOLUTION PRE-OPERATIVELY AND AN ESTIMATED 1.5 L INTRA-OPERATIVELY. ON POST-OPERATIVE DAY 1, THE PATIENT DEVELOPED ACUTE KIDNEY INJURY (AKI) AND WAS SUBSEQUENTLY ADMITTED FOR INPATIENT MANAGEMENT AND WAS DISCHARGED ON (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
841230 EDISON Focused ultrasound system for non-thermal, mechanical tissue ablation QGM HISTOSONICS, INC. 00850006962082

Patients

Seq Age Sex Outcome Treatment
1 41 YR Female Hospitalization