FARAWAVE 2.0 NAV

MDR report key: 24770213 · Received April 2, 2026

Report

Report Number
2124215-2026-17897
Event Type
Malfunction
Date Received
April 2, 2026
Date of Event
March 23, 2026
Report Date
April 1, 2026
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
QZI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. WE ARE UNABLE TO PROVIDE THE COMPLETE UNIQUE IDENTIFIER (UDI) AT THE TIME OF THE SUBMISSION OF THIS REPORT, AS IT WAS NOT YET AVAILABLE. IF ADDITIONAL DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A FARAWAVE CATHETER PRESENTED THE ERROR BADHDR AND WHEN IT WAS DISCONNECTED THE FLUSH PORT BROKE. A FARAWAVE 2.0 NAV WAS SELECTED FOR USE DURING A PULMONARY VEIN ISOLATION (PVI) PROCEDURE. THE ERROR BADHDR OCCURRED, HENCE THE CABLE WAS DISCONNECTED FROM CATHETER TO MSM, AND RECONNECTED, WHICH DID NOT SOLVE THE PROBLEM. THE ERROR CHANGED TO NO SYSTEM REFERENCE, THEN DISCONNECTED CABLE FROM CATHETER TO MSM AGAIN, TURNED OFF MSM, WAITED 5 SECONDS, TURNED MSM BACK ON AND THE CATHETER WAS RECONNECTED. IT WAS NEEDED TO REPLACED THE CATHETER. THEREFORE, IT WAS NOTE THAT SIDE PORT GOT DAMAGED AFTER CATHETER WAS REPLACED. LATER, IT WAS NOTED THAT THE FLUSHING PORT OF THE CATHETER BROKE WHEN DISCONNECTING AFTER ERROR OCCURRED. THERE WERE NO INFORMATION AVAILABLE INDICATING THE SIDE PORT WAS DEFECTIVE UPON OPENING. THE PROCEDURE WAS COMPLETED WITHOUT PATIENT COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
832305 FARAWAVE 2.0 NAV No Match QZI BOSTON SCIENTIFIC CORPORATION 0038255941

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown