PANEL PHOENIX PMIC-110
Report
- Report Number
- 1119779-2026-00602
- Event Type
- Malfunction
- Date Received
- April 2, 2026
- Date of Event
- March 11, 2026
- Report Date
- March 16, 2026
- Manufacturer
- BECTON DICKINSON & CO. (SPARKS)
- Product Code
- LON
- UDI-DI
- 30382904490363
- PMA / PMN Number
- SEE H11
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: K020322, K021954, K022172, K023273, K023301, K023568, K024152, K030091, K030677, K031306, K031679, K032131, K033784, K033889, K033907, K040006, K040106, K040716, K050089, K050555, K051689, K053241, K060214, K060217, K060218, K060493, K070809, K082538, K082852, K082913, K131331, AND K14046. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED WHILE USING THE PANEL PHOENIX PMIC-110 PATIENT ISOLATE (STAPHYLOCOCCUS SPP.) HAD A HIGH MIC (FALSE RESISTANT) RESULT FOR THE DRUG VANCOMYCIN BUT WHEN REPEATED THE RESULT WAS SENSITIVE. IT IS ALSO TO BE NOTED THAT THE UPSTREAM INSTRUMENT WAS FOUND TO BE CONTAMINATED. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 824235 | PANEL PHOENIX PMIC-110 | SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION | LON | BECTON DICKINSON & CO. (SPARKS) | 5325514 | 30382904490363 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |