VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 0°, AUTOCLAVABLE
Report
- Report Number
- 9610773-2026-02661
- Event Type
- Malfunction
- Date Received
- April 2, 2026
- Date of Event
- January 27, 2026
- Report Date
- April 2, 2026
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- HET
- UDI-DI
- 04042761074964
- PMA / PMN Number
- K190744
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE CAUSE OF THE FOREIGN MATERIAL IN THE LASER LIGHT CABLE (LLK) PLUG OF THE VIDEO CABLE SECTION COULD NOT BE IDENTIFIED, AND THE DAMAGE TO THE FIBER BONDING IN THE OUTER TUBE AREA (DISTAL END) WAS TRACED TO COMPONENT FAILURE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE RIGID VIDEO LAPAROSCOPE EXHIBITED FOREIGN MATERIAL IN THE LASER LIGHT CABLE (LLK) PLUG OF THE VIDEO-CABLE SECTION, AND DAMAGE TO THE FIBER BONDING IN THE OUTER TUBE AREA (DISTAL END). THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 823705 | VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 0°, AUTOCLAVABLE | RIGID VIDEO LAPAROSCOPE | HET | OLYMPUS WINTER & IBE GMBH | WA50040A | 04042761074964 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |