VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 0°, AUTOCLAVABLE

MDR report key: 24767450 · Received April 2, 2026

Report

Report Number
9610773-2026-02661
Event Type
Malfunction
Date Received
April 2, 2026
Date of Event
January 27, 2026
Report Date
April 2, 2026
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761074964
PMA / PMN Number
K190744
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE CAUSE OF THE FOREIGN MATERIAL IN THE LASER LIGHT CABLE (LLK) PLUG OF THE VIDEO CABLE SECTION COULD NOT BE IDENTIFIED, AND THE DAMAGE TO THE FIBER BONDING IN THE OUTER TUBE AREA (DISTAL END) WAS TRACED TO COMPONENT FAILURE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE RIGID VIDEO LAPAROSCOPE EXHIBITED FOREIGN MATERIAL IN THE LASER LIGHT CABLE (LLK) PLUG OF THE VIDEO-CABLE SECTION, AND DAMAGE TO THE FIBER BONDING IN THE OUTER TUBE AREA (DISTAL END). THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
823705 VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 0°, AUTOCLAVABLE RIGID VIDEO LAPAROSCOPE HET OLYMPUS WINTER & IBE GMBH WA50040A 04042761074964

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown