FDA Adverse Event
Malfunction
Summary report: N
AQUADIS 56
MDR report key: 24764089
·
Received April 2, 2026
Report
- Report Number
- 9616031-2026-0000003
- Event Type
- Malfunction
- Date Received
- April 2, 2026
- Date of Event
- March 13, 2026
- Report Date
- April 2, 2026
- Manufacturer
- GETINGE DISINFECTION AB
- Product Code
- MEC
- UDI-DI
- 07340153710351
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION.
Description of Event or Problem · 0
GETINGE DISINFECTION AB (GETINGE) RECEIVED INFORMATION OF SMOKE COMING FROM BOTTOM OF THE WASHER UNIT AQUADIS 56M. DURING TROUBLESHOOTING, IT WAS IDENTIFIED THAT THE CAPACITOR FOR THE DRAIN MOTOR HAD SHORTED AT THE TERMINAL. A REPLACEMENT PART WAS REQUIRED AND ORDERED, RESULTING IN A RETURN VISIT TO INSTALL A NEW CAPACITOR. SUBSEQUENTLY, THE DRAIN PUMP WAS REPLACED, AND THE UNIT IS NOW OPERATING AS DESIGNED AND HAS SUCCESSFULLY PASSED ALL FUNCTIONAL TESTING. MANUFACTURER HAS REQUESTED THE RETURN OF THE PART FOR FURTHER ANALYSIS. THERE WAS NO INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 824893 | AQUADIS 56 | DISINFECTOR, MEDICAL DEVICES | MEC | GETINGE DISINFECTION AB | 56M | 07340153710351 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |