FDA Adverse Event Malfunction Summary report: N

AQUADIS 56

MDR report key: 24764089 · Received April 2, 2026

Report

Report Number
9616031-2026-0000003
Event Type
Malfunction
Date Received
April 2, 2026
Date of Event
March 13, 2026
Report Date
April 2, 2026
Manufacturer
GETINGE DISINFECTION AB
Product Code
MEC
UDI-DI
07340153710351
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION.

Description of Event or Problem · 0

GETINGE DISINFECTION AB (GETINGE) RECEIVED INFORMATION OF SMOKE COMING FROM BOTTOM OF THE WASHER UNIT AQUADIS 56M. DURING TROUBLESHOOTING, IT WAS IDENTIFIED THAT THE CAPACITOR FOR THE DRAIN MOTOR HAD SHORTED AT THE TERMINAL. A REPLACEMENT PART WAS REQUIRED AND ORDERED, RESULTING IN A RETURN VISIT TO INSTALL A NEW CAPACITOR. SUBSEQUENTLY, THE DRAIN PUMP WAS REPLACED, AND THE UNIT IS NOW OPERATING AS DESIGNED AND HAS SUCCESSFULLY PASSED ALL FUNCTIONAL TESTING. MANUFACTURER HAS REQUESTED THE RETURN OF THE PART FOR FURTHER ANALYSIS. THERE WAS NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
824893 AQUADIS 56 DISINFECTOR, MEDICAL DEVICES MEC GETINGE DISINFECTION AB 56M 07340153710351

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown