ALLURION DEVICE
Report
- Report Number
- 3011299930-2026-00002
- Event Type
- Malfunction
- Date Received
- April 2, 2026
- Date of Event
- March 5, 2026
- Report Date
- April 2, 2026
- Manufacturer
- ALLURION TECHNOLOGIES, INC.
- Product Code
- LTI
- UDI-DI
- 00810010540317
- PMA / PMN Number
- P250023
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
HYPERINFLATION OF THE DEVICE MAY BE ASSOCIATED WITH THE ISOLATION OF LACTOBACILLUS ACIDOPHILUS/GASSERI, WHICH WAS IDENTIFIED FOLLOWING CULTURING OF THE FLUID FROM INSIDE THE INTRAGASTRIC BALLOON DURING THE ENDOSCOPIC REMOVAL. THESE FERMENTATIVE ORGANISMS ARE KNOWN TO PRODUCE GAS AS A METABOLIC BYPRODUCT, WHICH MAY INCREASE INTERNAL PRESSURE AND CONTRIBUTE TO VOLUMETRIC EXPANSION OF THE DEVICE. THIS FINDING MAY PROVIDE A POTENTIAL EXPLANATION FOR THE OBSERVED DEVICE HYPERINFLATION. HOWEVER, LACTOBACILLUS SPECIES ARE PART OF THE NORMAL GASTROINTESTINAL FLORA, AND THE SOURCE OF THE ORGANISM COULD NOT BE CONFIRMED, INCLUDING THE POSSIBILITY OF EXTERNAL CONTAMINATION. THE DEVICE IS AVAILABLE FOR RETURN TO THE MANUFACTURER AND HAS BEEN REQUESTED. FURTHER EVALUATION IS PENDING THE RETURN AND ANALYSIS OF THE DEVICE.
THE PATIENT UNDERWENT PLACEMENT OF THE ALLURION BALLOON ON (B)(6) 2026. THE PATIENT SUBSEQUENTLY REPORTED PERSISTENT ABDOMINAL PAIN AND COMPLETE INTOLERANCE TO ORAL INTAKE. AN ABDOMINAL X-RAY PERFORMED ON (B)(6) 2026 WHICH SHOWED THAT THE DEVICE WAS HYPERINFLATED. THE DEVICE WAS ENDOSCOPICALLY PUNCTURED AND REMOVED ON (B)(6) 2026. NO ADDITIONAL COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 829825 | ALLURION DEVICE | GASTRIC BALLOON | LTI | ALLURION TECHNOLOGIES, INC. | 10D-CE | 309191-1 | 00810010540317 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Female | Required Intervention |