FDA Adverse Event Malfunction Summary report: N

ENDOWRIST SP

MDR report key: 24759632 · Received April 1, 2026

Report

Report Number
2955842-2026-20163
Event Type
Malfunction
Date Received
April 1, 2026
Date of Event
March 4, 2026
Report Date
April 1, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874114285
PMA / PMN Number
K182371
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE SINGLE PORT (SP) MONOPOLAR CURVED SCISSORS (MCS) WAS ANALYZED AND FOUND TO HAVE A BROKEN INSTRUMENT TUBE ADAPTER. THIS COMPONENT IS LOCATED AT THE DISTAL END OF THE INSTRUMENT AND SERVES AS THE INTERFACE BETWEEN THE INSTRUMENT AND THE USER INSTALLED TIP ACCESSORY. THE BROKEN PIECE WAS NOT RETURNED AND MEASURES APPROXIMATELY 1.10MM X 2.10MM IN SIZE. THE COMPLAINT REGARDING INSTRUMENT TIP WAS DAMAGED WAS CONFIRMED BY FAILURE ANALYSIS. THE PROBABLE ROOT CAUSE OF A DAMAGED TUBE ADAPTER IS ATTRIBUTED TO EITHER TIP ACCESSORY INSTALLATION ISSUES OR DAMAGE DURING USE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE SINGLE PORT (SP) MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT TIP WAS DAMAGED, AND THE SCISSOR TIP WAS UNABLE TO BE ATTACHED AND USED PROPERLY. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470811 ENDOWRIST SP MONOPOLAR CURVED SCISSORS NAY INTUITIVE SURGICAL, INC 430004-57 U10230824 0003 00886874114285

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES