ARCHITECT FREE T4
Report
- Report Number
- 3005094123-2026-00155
- Event Type
- Malfunction
- Date Received
- April 1, 2026
- Date of Event
- March 27, 2026
- Report Date
- April 8, 2026
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- CEC
- UDI-DI
- 00380740173685
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 07K65-78 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 7K65, WITH 510K/PMA/BLA NUMBER K173122.
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. UPDATED SECTION B5 WITH ADDITIONAL PATIENT INFORMATION.
THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT FREE T4 RESULTS GENERATED ON THE ARCHITECT I2000 ANALYZER FOR 2 PATIENTS. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE: 11-17.7 PMOL/L): TESTING WAS COMPLETED ON (B)(6) 2026: SID (B)(6) (FEMALE, 30 YEARS-OLD): INITIAL RESULT = 33.23 PMOL/L; REPEAT RESULT = 10.74 PMOL/L SID (B)(6) (FEMALE, 31 YEARS-OLD): INITIAL RESULT = 18.01 PMOL/L; REPEAT RESULT = 13.8 PMOL/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
ON (B)(6) 2026, THE CUSTOMER PROVIDED AN ADDITIONAL PATIENT WITH A FALSELY ELEVATED ARCHITECT FREE T4 (FT4) RESULT. THE PATIENT SAMPLE GENERATED AN INITIAL FT4 RESULT OF 30.18 PMOL/L AND THE REPEAT RESULT OF 13.59 PMOL/L. THE PATIENT IS A 33-YEAR-OLD FEMALE. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 78552 | ARCHITECT FREE T4 | RADIOIMMUNOASSAY, FREE THYROXINE | CEC | ABBOTT IRELAND DIAGNOSTICS DIVISION | 77458UD03 | 00380740173685 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |