FDA Adverse Event Injury Summary report: N

NBCA LIQUID EMBOLIC KIT

MDR report key: 2475740 · Received March 2, 2012

Report

Report Number
1058196-2012-00102
Event Type
Injury
Date Received
March 2, 2012
Date of Event
February 2, 2012
Report Date
February 20, 2012
Manufacturer
CORDIS NEUROVASCULAR, INC.
Product Code
KGG
PMA / PMN Number
P990040
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ONE PATIENT HAD FOUR EMBOLIZATIONS OVER A 5-MONTH PERIOD, ONE HAD THREE EMBOLIZATIONS OVER 3 MONTHS COMPLICATED BY HEMORRHAGE AFTER EMBOLIZATION REQUIRING EMERGENT EVACUATION OF THE HEMATOMA, AND THE THIRD PATIENT HAD A SINGLE EMBOLIZATION. IN ALL THREE PATIENTS, SURGICAL AND AUTOPSY SPECIMENS SHOWED AN INFLAMMATORY RESPONSE WITHIN THE EMBOLIZED VASCULATURE WITH A PROMINENT EOSINOPHILIC INFILTRATE. CONCLUSION: THE EOSINOPHILIC VASCULITIS SEEN IN THE PATHOLOGY SPECIMENS MAY REPRESENT A PREVIOUSLY UNDOCUMENTED HYPERSENSITIVITY REACTION FOLLOWING EXPOSURE TO N-BCA, WITH THE POTENTIAL FOR ADVERSE SEQUELAE, INCLUDING INCREASED RISK OF HEMORRHAGE AS WAS SEEN IN ONE OF OUR PATIENTS. JOURNAL OF NEUROINTERVENTIONAL SURGERY MARCH 2011 3 (1) (PP 21-24) 'VASCULAR INFLAMMATION WITH EOSINOPHILS AFTER THE USE OF N-BUTYL CYANOACRYLATE LIQUID EMBOLIC SYSTEM', QUINN J.C., MITTAL N., BAISRE A., CHO E.S., SHARER L.R., CANDHI C., PRESTIGIACOMA C.J. REPORTED THREE CASES OF RESTRICTED EOSINOPHILIC VASCULITIS SECONDARY TO NBCA SENSITIZATION. ONE OF THESE CASES WAS COMPLICATED BY HEMORRHAGE AFTER EMBOLIZATION. ALL PATIENTS PRESENTED WITH HEADACHE. CASE 1 REPORTS A WOMAN INITIALLY PRESENTING WITH NEW-ONSET SEIZURES WAS DIAGNOSED WITH A 3 CM LEFT FRONTAL AVM (SPETZLEREMARTIN II). THE PATIENT UNDERWENT FOUR STAGES OF ENDOVASCULAR EMBOLIZATION WITH N-BCA OVER THE COURSE OF THE NEXT 5 MONTHS. A TOTAL OF SIX PEDICLES WERE EMBOLIZED OVER THE FOUR STAGES WITH N-BCA CONCENTRATIONS RANGING FROM 20 TO 33% IN LIPIODOL. ONE MONTH AFTER THE FINAL STAGE OF EMBOLIZATION, A LEFT FRONTAL CRANIOTOMY AND TOTAL RESECTION OF THE AVM WAS PERFORMED. INTRAOPERATIVELY, ARTERIAL PEDICLES AND GLUE CASTS WERE IDENTIFIED AND RESECTED. POSTOPERATIVE ANGIOGRAPHY CONFIRMED THE COMPLETE RESECTION OF THE LESION. THE PATIENT'S POSTOPERATIVE COURSE WAS UNEVENTFUL AND SHE WAS DISCHARGED ON POSTOPERATIVE DAY 5 WITH NO RESIDUAL NEUROLOGIC DEFICITS. MICROSCOPIC EXAMINATION OF THE SURGICAL SPECIMEN DEMONSTRATED THE EXPECTED HYALINIZED AVM VASCULATURE WITH INTRALUMINAL N-BCA. THE VESSEL WALLS CONTAINED A DENSE GRANULOCYTIC REACTION WITH A PROMINENT EOSINOPHILIC INFILTRATE SUGGESTIVE OF VASCULITIS. ACCUMULATION OF FOREIGN-BODY GIANT CELLS AND HEMOSIDERIN-LADEN MACROPHAGES WAS ALSO NOTED. THERE WAS NO EVIDENCE OF EXTRAVASATION OF THE EMBOLIC AGENT. ROUTINE BLOOD WORK SHOWED NO PERIPHERAL EOSINOPHILIA OR SIGNS OF SYSTEMIC INVOLVEMENT. THE 158 PATIENTS WITH VARYING GRADES OF AVMS WERE TREATED WITH N-BCA EMBOLIZATION, FOLLOWED BY OPEN RESECTION BETWEEN 2002 AND 2008. ALL SPECIMENS OBTAINED DURING RESECTION WERE SENT FOR PATHOLOGIC REVIEW. HISTOPATHOLOGIC EXAMINATION OF THE AVMS DEMONSTRATED THE HISTOPATHOLOGIC PROGRESSION AS HAS BEEN PREVIOUSLY DESCRIBED. UNIQUE TO THIS SERIES OF THREE PATIENTS AMONG THE 158 CASES REVIEWED WAS THE PRESENCE OF A PROMINENT EOSINOPHILIC INFILTRATE AFFECTING THE ENDOTHELIUM AND EXTENDING INTO THE MEDIA OF THE EMBOLIZED VESSELS, SUGGESTING A SEPARATE INFLAMMATORY/IMMUNOLOGICAL PROCESS IN ADDITION TO THE NATURAL PROGRESSION WHICH HAS BEEN PREVIOUSLY DESCRIBED. OTHER PORTIONS OF THE NON-EMBOLIZED AVM DID SHOW A SIMILAR EOSINOPHILIC INFILTRATE, ARGUING THAT THIS IS PROBABLY A NON-FOCAL REACTION. IN ALL CASES THERE WAS NO EVIDENCE OF PERIPHERAL EOSINOPHILIA OR ANY CLINICAL SIGNS OF AN UNDERLYING SYSTEMIC CAUSE FOR A CNS RESTRICTED EOSINOPHILIC INFILTRATE, ALTHOUGH DEDICATED SEROLOGIC TESTING WAS NOT DONE IN THESE PATIENTS. THE PRESENCE OF EOSINOPHILS IN THE CHRONIC INFLAMMATORY MILIEU PROBABLY REPRESENTS A RESTRICTED EOSINOPHILIC VASCULITIS CAUSED BY HOST SENSITIZATION TO N-BCA EPITOPES. THE ARTICLE STATES THAT THE CLINICAL CONSEQUENCES OF THE REPORTED INFLAMMATORY/IMMUNOLOGICAL RESPONSE ARE DIFFICULT TO ASSESS IN THE SERIES FOR A NUMBER OF REASONS, INCLUDING THE PATIENT HAD BASELINE NEUROLOGICAL DEFICITS FROM THEIR PRIMARY VASCULAR MALFORMATIONS, THE EXPOSURE TIME TO NBCA WAS RELATIVELY SHORT, AND FOR TWO PATIENTS, ALL EMBOLIZED VASCULATURE WAS RESECTED DURING THE STATED SURGICAL PROCEDURE. IT FURTHER REPORTS THAT BECAUSE THIS REACTION WAS DESCRIBED IN PATIENTS IN WHOM A STAGED RESECTION WAS PLANNED, IT IS ENTIRELY POSSIBLE THAT THE CLINICAL SEQUEL OF N-BCA EXPOSURE DID NOT HAVE TIME TO MANIFEST ITSELF. IT IS REPORTED THAT THE IMPLICATIONS OF THESE FINDINGS MAY BE MORE RELEVANT IN CASES WHERE N-BCA EMBOLIZATION IS TO BE USED AS A MONOTHERAPY OR IN CASES WHERE NOT ALL THE EMBOLIC AGENT IS FULLY RESECTED. THE AUTHORS HYPOTHESIZE THAT THE THREE CASES OF RESTRICTED EOSINOPHILIC VASCULITIS ARE SECONDARY TO N-BCA SENSITIZATION AND THAT THEY BELIEVE THIS MAY REPRESENT A RARE BUT SIGNIFICANT REACTION IN SOME PATIENTS HAS THE POTENTIAL TO LEAD TO INCREASED RUPTURE RISK AND, POTENTIALLY, OTHER NEUROLOGICAL SEQUELAE. THE PRODUCT IS NOT AVAILABLE FOR EVALUATION AND TESTING. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS UPON RECEIPT. THIS IS ONE OF FOUR PRODUCTS REPORTED IN THE SAME JOURNAL ARTICLE/LITERATURE SEARCH. PLEASE REFERENCE MFR. REPORT # 1058196-2012-00101; # 1058196-2012-00102; # 1058196-2012-00103; AND # 1058196-2012-00104.

Additional Manufacturer Narrative · 1

PLEASE NOTE THAT THE PRODUCT CATALOG NUMBER WAS CORRECTED. JOURNAL OF NEUROINTERVENTIONAL SURGERY MARCH 2011 3 (1) (PP 21-24) "VASCULAR INFLAMMATION WITH EOSINOPHILS AFTER THE USE OF N-BUTYL CYANOACRYLATE LIQUID EMBOLIC SYSTEM", QUINN J.C., MITTAL N., BAISRE A., CHO E.S., SHARER L.R., CANDHI C., PRESTIGIACOMA C.J. REPORTED THREE CASES OF RESTRICTED EOSINOPHILIC VASCULITIS SECONDARY TO NBCA SENSITIZATION. IN ONE CASE A WOMAN INITIALLY PRESENTING WITH NEW-ONSET SEIZURES WAS DIAGNOSED WITH A 3 CM LEFT FRONTAL AVM (SPETZLEREMARTIN II). THE PATIENT UNDERWENT FOUR STAGES OF ENDOVASCULAR EMBOLIZATION WITH N-BCA OVER THE COURSE OF THE NEXT 5 MONTHS. A TOTAL OF SIX PEDICLES WERE EMBOLIZED OVER THE FOUR STAGES WITH N-BCA CONCENTRATIONS RANGING FROM 20 TO 33% IN LIPIODOL. ONE MONTH AFTER THE FINAL STAGE OF EMBOLIZATION, A LEFT FRONTAL CRANIOTOMY AND TOTAL RESECTION OF THE AVM WAS PERFORMED. INTRAOPERATIVELY, ARTERIAL PEDICLES AND GLUE CASTS WERE IDENTIFIED AND RESECTED. POSTOPERATIVE ANGIOGRAPHY CONFIRMED THE COMPLETE RESECTION OF THE LESION. THE PATIENT'S POSTOPERATIVE COURSE WAS UNEVENTFUL AND SHE WAS DISCHARGED ON POSTOPERATIVE DAY 5 WITH NO RESIDUAL NEUROLOGIC DEFICITS. MICROSCOPIC EXAMINATION OF THE SURGICAL SPECIMEN DEMONSTRATED THE EXPECTED HYALINIZED AVM VASCULATURE WITH INTRALUMINAL N-BCA. THE VESSEL WALLS CONTAINED A DENSE GRANULOCYTIC REACTION WITH A PROMINENT EOSINOPHILIC INFILTRATE SUGGESTIVE OF VASCULITIS. ACCUMULATION OF FOREIGN-BODY GIANT CELLS AND HEMOSIDERIN-LADEN MACROPHAGES WAS ALSO NOTED. THERE WAS NO EVIDENCE OF EXTRAVASATION OF THE EMBOLIC AGENT. ROUTINE BLOOD WORK SHOWED NO PERIPHERAL EOSINOPHILIA OR SIGNS OF SYSTEMIC INVOLVEMENT; 158 PATIENTS WITH VARYING GRADES OF AVMS WERE TREATED WITH N-BCA EMBOLIZATION, FOLLOWED BY OPEN RESECTION BETWEEN 2002 AND 2008. ALL SPECIMENS OBTAINED DURING RESECTION WERE SENT FOR PATHOLOGIC REVIEW. HISTOPATHOLOGIC EXAMINATION OF THE AVMS DEMONSTRATED THE HISTOPATHOLOGIC PROGRESSION AS HAS BEEN PREVIOUSLY DESCRIBED. UNIQUE TO THIS SERIES OF THREE PATIENTS AMONG THE 158 CASES REVIEWED WAS THE PRESENCE OF A PROMINENT EOSINOPHILIC INFILTRATE AFFECTING THE ENDOTHELIUM AND EXTENDING INTO THE MEDIA OF THE EMBOLIZED VESSELS, SUGGESTING A SEPARATE INFLAMMATORY/IMMUNOLOGICAL PROCESS IN ADDITION TO THE NATURAL PROGRESSION WHICH HAS BEEN PREVIOUSLY DESCRIBED. OTHER PORTIONS OF THE NON-EMBOLIZED AVM DID SHOW A SIMILAR EOSINOPHILIC INFILTRATE, ARGUING THAT THIS IS PROBABLY A NON-FOCAL REACTION. IN ALL CASES THERE WAS NO EVIDENCE OF PERIPHERAL EOSINOPHILIA OR ANY CLINICAL SIGNS OF AN UNDERLYING SYSTEMIC CAUSE FOR A CNS RESTRICTED EOSINOPHILIC INFILTRATE, ALTHOUGH DEDICATED SEROLOGIC TESTING WAS NOT DONE IN THESE PATIENTS. THE PRESENCE OF EOSINOPHILS IN THE CHRONIC INFLAMMATORY MILIEU PROBABLY REPRESENTS A RESTRICTED EOSINOPHILIC VASCULITIS CAUSED BY HOST SENSITIZATION TO N-BCA EPITOPES. THE ARTICLE STATES THAT THE CLINICAL CONSEQUENCES OF THE REPORTED INFLAMMATORY/IMMUNOLOGICAL RESPONSE ARE DIFFICULT TO ASSESS IN THE SERIES FOR A NUMBER OF REASONS, INCLUDING THE PATIENT HAD BASELINE NEUROLOGICAL DEFICITS FROM THEIR PRIMARY VASCULAR MALFORMATIONS, THE EXPOSURE TIME TO NBCA WAS RELATIVELY SHORT, AND FOR TWO PATIENTS, ALL EMBOLIZED VASCULATURE WAS RESECTED DURING THE STATED SURGICAL PROCEDURE. IT FURTHER REPORTS THAT BECAUSE THIS REACTION WAS DESCRIBED IN PATIENTS IN WHOM A STAGED RESECTION WAS PLANNED, IT IS ENTIRELY POSSIBLE THAT THE CLINICAL SEQUEL OF N-BCA EXPOSURE DID NOT HAVE TIME TO MANIFEST ITSELF. IT IS REPORTED THAT THE IMPLICATIONS OF THESE FINDINGS MAY BE MORE RELEVANT IN CASES WHERE N-BCA EMBOLIZATION IS TO BE USED AS A MONOTHERAPY OR IN CASES WHERE NOT ALL THE EMBOLIC AGENT IS FULLY RESECTED. THE AUTHORS HYPOTHESIZE THAT THE THREE CASES OF RESTRICTED EOSINOPHILIC VASCULITIS ARE SECONDARY TO N-BCA SENSITIZATION AND THAT THEY BELIEVE THIS MAY REPRESENT A RARE BUT SIGNIFICANT REACTION IN SOME PATIENTS HAS THE POTENTIAL TO LEAD TO INCREASED RUPTURE RISK AND, POTENTIALLY, OTHER NEUROLOGICAL SEQUELAE. THE PRODUCT WAS NOT RETURNED FOR INSPECTION. THE LOT NUMBER OF THE TRUFILL NBCA USED IN THE PROCEDURE IS NOT KNOWN; THEREFORE A DEVICE HISTORY RECORD REVIEW CANNOT BE COMPLETED. THE AUTHORS REPORT THAT THE IMPLICATIONS OF THE FINDINGS OF EOSINOPHILIC VASCULITIS MAY BE MORE RELEVANT IN CASES WHERE N-BCA EMBOLIZATIONS IS TO BE USED AS A MONOTHERAPY OR IN CASES WHERE NOT ALL THE EMBOLIC AGENT IS FULLY RESECTED. THE INSTRUCTIONS FOR USE OUTLINES THAT TRUFILL NBCA LIQUID EMBOLIC SYSTEM IS INDICATED FOR THE EMBOLIZATION OF CEREBRAL AVMS WHEN PRESURGICAL DEVASCULARIZATION IS DESIRED. IT FURTHER WARNS THAT THE SAFETY AND EFFICACY OF THIS PRODUCT AS A LONG-TERM IMPLANT HAS NOT BEEN ESTABLISHED. IT IS CONTRAINDICATED FOR USE WITH PREVIOUS HISTORY OF REACTIONS TO COMPONENTS. THERE ARE NO INDICATIONS OF ANY DEVICE MANUFACTURING ISSUES RELATED TO THE EVENT. INFECTION/INFLAMMATION ARE OUTLINED IN THE INSTRUCTIONS FOR USE AS ADVERSE EVENTS THAT MAY BE ASSOCIATED WITH EMBOLIZATION PROCEDURES (INCLUDING THOSE OBSERVED DURING THE CLINICAL STUDY), AND MAY OCCUR AT ANY TIME DURING OR AFTER THE PROCEDURE. THEREFORE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME.

Description of Event or Problem · 1

DURING A LITERATURE SEARCH THE FOLLOWING ARTICLE WAS FOUND: JOURNAL OF NEUROINTERVENTIONAL SURGERY MARCH 2011 3 (1) (PP 21-24) 'VASCULAR INFLAMMATION WITH EOSINOPHILS AFTER THE USE OF N-BUTYL CYANOACRYLATE LIQUID EMBOLIC SYSTEM', QUINN J.C., MITTAL N., BAISRE A., CHO E.S., SHARER L.R., CANDHI C., PRESTIGIACOMA C.J. THE ARTICLE REPORTED THREE CASES OF RESTRICTED EOSINOPHILIC VASCULITIS SECONDARY TO NBCA SENSITIZATION. THE CASES OF THREE PATIENTS ARE REPORTED WHO UNDERWENT ENDOVASCULAR EMBOLIZATION WITH N-BCA, FOLLOWED BY RESECTION IN TWO AND POST-EMBOLIZATION HEMORRHAGE WITH EMERGENT EVACUATION IN ONE, WITH HISTOLOGIC DEMONSTRATION OF AN EOSINOPHILIC VASCULITIS FOUND IN RESECTED AVM SPECIMENS. THIS IS PROBABLY THE FIRST REPORT OF THIS TISSUE REACTION, WHICH MAY HAVE THEORETICALLY SERIOUS CLINICAL IMPLICATIONS. CLINICAL PRESENTATION: IN THIS SERIES, THREE PATIENTS (2 WOMEN, 1 MAN) PRESENTED WITH INTRACRANIAL AVMS (SPETZLER-MARTIN I-III) WITH THE LESIONS LOCATED IN THE FRONTAL LOBE IN TWO OF THE PATIENTS AND IN THE PARIETAL LOBE IN ONE. ALL PATIENTS PRESENTED WITH HEADACHE, AND ONE ALSO HAD NEW-ONSET SEIZURES. INTERVENTION: ALL PATIENTS UNDERWENT EMBOLIZATION WITH N-BCA BEFORE A PLANNED, STAGED SURGICAL RESECTION OF THE EMBOLIZED AVMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NBCA LIQUID EMBOLIC KIT CNV_NBCA (MFE) KGG CORDIS NEUROVASCULAR, INC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1