FDA Adverse Event Injury Summary report: N

INX FOR INLINE

MDR report key: 24756717 · Received April 1, 2026

Report

Report Number
1523574-2026-00026
Event Type
Injury
Date Received
April 1, 2026
Date of Event
March 1, 2026
Report Date
April 16, 2026
Manufacturer
FERNO-WASHINGTON, INC.
Product Code
FPO
UDI-DI
00190790001339
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE SUBJECT STRETCHER WAS RETURNED TO MANUFACTURER FOR A VISUAL AND FUNCTIONAL EVALUATION. IT WAS CONFIRMED A MOUNTING BLOCK HAD BROKEN WHICH RESULTED IN THE UNINTENDED MOVEMENT OF THE LEG ASSEMBLY THE STRETCHER WAS REPAIRED AND RETURNED TO SERVICE.

Description of Event or Problem · 0

THE END USER REPORTED WHILE UNLOADING A PATIENT FROM THE AMBULANCE ON THE STRETCHER, AFTER THE LEGS WERE EXTENDED TO THE GROUND AND THE LOAD END OPERATOR RELEASED THE SAFETY BAIL, ONE OF THE STRETCHER LEGS "BROKE" RESULTING IN UNINTENDED MOVEMENT OF THE STRETCHER. THE PATIENT WAS NOT INJURED; HOWEVER, THE MEDIC'S ARM WAS PINNED BETWEEN THE STRETCHER AND THE BACK OF THE AMBULANCE. THE MEDIC SOUGHT MEDICAL EVALUATION AND THE RESULTING INJURY WAS A BRUISE AND THE MEDIC WAS RELEASED BACK TO WORK. AN EVALUATION OF THE STRETCHER HAS BEEN INITIATED.

Description of Event or Problem · 0

THE END USER REPORTED WHILE UNLOADING A PATIENT FROM THE AMBULANCE ON THE STRETCHER, AFTER THE LEGS WERE EXTENDED TO THE GROUND AND THE LOAD END OPERATOR RELEASED THE SAFETY BAIL, ONE OF THE STRETCHER LEGS "BROKE" RESULTING IN UNINTENDED MOVEMENT OF THE STRETCHER. THE PATIENT WAS NOT INJURED; HOWEVER, THE MEDIC'S ARM WAS PINNED BETWEEN THE STRETCHER AND THE BACK OF THE AMBULANCE. THE MEDIC SOUGHT MEDICAL EVALUATION AND THE RESULTING INJURY WAS A BRUISE AND THE MEDIC WAS RELEASED BACK TO WORK. AN EVALUATION OF THE STRETCHER HAS BEEN INITIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470784 INX FOR INLINE INX FOR INLINE FPO FERNO-WASHINGTON, INC. 0015811 M991690 00190790001339

Patients

Seq Age Sex Outcome Treatment
1