Report

Report Number
1710034-2026-00360
Event Type
Malfunction
Date Received
April 1, 2026
Date of Event
February 26, 2026
Report Date
April 2, 2026
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825448
PMA / PMN Number
SEE H.11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
RISK MANAGER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 382544 AND LOT NUMBER 5233462. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE, AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT SIMILAR ISSUES DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

G.4. K201075; K251654. H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO NEW INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE NEEDLE DID NOT RETRACT. 18G IV NEEDLE SAFETY MALFUNCTIONED. NEEDLE SAFETY WOULD NOT ENGAGE DESPITE PRESSING THE SAFETY BUTTON MULTIPLE TIMES. ADDITIONAL INFORMATION RECEIVED 3/26/2026: DATE OF EVENT: 2/26/2026 NO PATIENT HARM, INJURY, OR COMPLICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
817236 INSYTE AUTOGUARD PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 5233462 00382903825448

Patients

Seq Age Sex Outcome Treatment
1 78 YR Male