FDA Adverse Event Malfunction Summary report: N

COBAS E 411 ANALYZER (RACK SYSTEM)

MDR report key: 24753443 · Received April 1, 2026

Report

Report Number
1823260-2026-01200
Event Type
Malfunction
Date Received
April 1, 2026
Date of Event
March 11, 2026
Report Date
April 1, 2026
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
UDI-DI
04015630924721
PMA / PMN Number
K961481
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE TSH REAGENT LOT NUMBER WAS 891754. THE FT4 REAGENT LOT NUMBER WAS 883869. THE TSH AND THE FT4 REAGENTS' EXPIRATION DATES WERE NOT PROVIDED. THE FIELD SERVICE ENGINEER (FSE) FOUND AN ISSUE WITH THE SAMPLE PROBE TUBING. HE REPLACED THE PINCH VALVE TUBINGS, THE SAMPLE PROBE TUBINGS, AND PRIMED THE SAMPLE/REAGENT PROBE. HE ALSO PRIMED THE SIPPER, ADJUSTED THE SAMPLE PROBE LIQUID LEVEL DETECTION VOLTAGE, AND PERFORMED MAINTENANCE ON THE INSTRUMENT. QC WAS PERFORMED, AND IT WAS WITHIN SPECIFICATIONS. THE FSE VERIFIED THAT THE INSTRUMENT WAS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER ISSUES WERE REPORTED AFTER THE SERVICE VISIT. THE CAUSE OF THE EVENT WAS DUE TO A COMPONENT FAILURE (AN ISSUE WITH THE SAMPLE PROBE TUBING).

Description of Event or Problem · 0

WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH ELECSYS TSH ASSAY AND ELECSYS FT4 G4 ASSAY ON A COBAS E 411 ANALYZER (RACK SYSTEM). TSH: INITIAL RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 1ST REPEAT RESULT: 0.005 UIU/ML (ACCOMPANIED BY A DATA FLAG). 2ND REPEAT RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 3RD REPEAT RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 4TH REPEAT RESULT: 0.005 UIU/ML (ACCOMPANIED BY A DATA FLAG). 5TH REPEAT RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 6TH REPEAT RESULT: 1.6 UIU/ML (TESTED IN A DIFFERENT LABORATORY). FT4: INITIAL RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 1ST REPEAT RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 2ND REPEAT RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 3RD REPEAT RESULT: "NO VALUE" WAS GENERATED AND (ACCOMPANIED BY A "<SIGNAL" DATA FLAG). 4TH REPEAT RESULT: 16 PMOL/L (TESTED IN A DIFFERENT LABORATORY).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
18730 COBAS E 411 ANALYZER (RACK SYSTEM) IMMUNOCHEMISTRY ANALYZER JJE ROCHE DIAGNOSTICS E 411 04015630924721

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown