COBAS® EGFR MUTATION TEST V2 P150044
Report
- Report Number
- 2243471-2026-01198
- Event Type
- Malfunction
- Date Received
- April 1, 2026
- Date of Event
- March 13, 2026
- Report Date
- April 30, 2026
- Manufacturer
- ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
- Product Code
- OWD
- UDI-DI
- 00875197005448
- PMA / PMN Number
- P150044
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE DATA SHOWS SIGMOIDAL CURVES AND IS CONSISTENT WITH ALL THREE RUNS. NO ABNORMALITIES WERE OBSERVED IN ANY OF THE CONTROLS. OVERALL, THE RESULTS APPEAR TO BE TRUE POSITIVE RESULTS. THE OBSERVED DISCREPANCY MAY BE DUE TO DIFFERENCES IN TECHNOLOGY, SENSITIVITY, AND/OR SPECIFICITY BETWEEN THE ASSAYS. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.
THE SERIAL NUMBER OF THE ANALYZER IS (B)(4). THE INVESTIGATION IS ONGOING. FULL E1 INITIAL REPORTER ESTABLISHMENT NAME:(B)(6).
THERE WAS AN ALLEGATION OF A QUESTIONABLE COBAS® EGFR MUTATION TEST V2 ASSAY RESULT FOR A PATIENT SAMPLE TESTED ON THE COBAS Z480 ANALYZER. (B)(6) 2026: EGFR EXON20INS - MUTATION DETECTED (B)(6) 2026: EGFR EXON20INS - MUTATION DETECTED (B)(6) 2026: EGFR EXON20INS - MUTATION DETECTED THE SAMPLE WAS REPEATED ON TWO OTHER METHODS, AMOY AND IDYLLA, AND NO MUTATION WAS DETECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 18727 | COBAS® EGFR MUTATION TEST V2 P150044 | SOMATIC GENE MUTATION DETECTION SYSTEM | OWD | ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG | M25692 | 00875197005448 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |