FDA Adverse Event Malfunction Summary report: N

COBAS® EGFR MUTATION TEST V2 P150044

MDR report key: 24753436 · Received April 1, 2026

Report

Report Number
2243471-2026-01198
Event Type
Malfunction
Date Received
April 1, 2026
Date of Event
March 13, 2026
Report Date
April 30, 2026
Manufacturer
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
Product Code
OWD
UDI-DI
00875197005448
PMA / PMN Number
P150044
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DATA SHOWS SIGMOIDAL CURVES AND IS CONSISTENT WITH ALL THREE RUNS. NO ABNORMALITIES WERE OBSERVED IN ANY OF THE CONTROLS. OVERALL, THE RESULTS APPEAR TO BE TRUE POSITIVE RESULTS. THE OBSERVED DISCREPANCY MAY BE DUE TO DIFFERENCES IN TECHNOLOGY, SENSITIVITY, AND/OR SPECIFICITY BETWEEN THE ASSAYS. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.

Additional Manufacturer Narrative · 0

THE SERIAL NUMBER OF THE ANALYZER IS (B)(4). THE INVESTIGATION IS ONGOING. FULL E1 INITIAL REPORTER ESTABLISHMENT NAME:(B)(6).

Description of Event or Problem · 0

THERE WAS AN ALLEGATION OF A QUESTIONABLE COBAS® EGFR MUTATION TEST V2 ASSAY RESULT FOR A PATIENT SAMPLE TESTED ON THE COBAS Z480 ANALYZER. (B)(6) 2026: EGFR EXON20INS - MUTATION DETECTED (B)(6) 2026: EGFR EXON20INS - MUTATION DETECTED (B)(6) 2026: EGFR EXON20INS - MUTATION DETECTED THE SAMPLE WAS REPEATED ON TWO OTHER METHODS, AMOY AND IDYLLA, AND NO MUTATION WAS DETECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
18727 COBAS® EGFR MUTATION TEST V2 P150044 SOMATIC GENE MUTATION DETECTION SYSTEM OWD ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG M25692 00875197005448

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown