UNICEL DXC 600 SYNCHRON SYSTEM
Report
- Report Number
- 2050012-2012-00628
- Event Type
- Malfunction
- Date Received
- March 2, 2012
- Date of Event
- February 14, 2012
- Report Date
- February 14, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
CUSTOMER HAS NOT CALLED BACK TO REPORT ANY FURTHER ISSUES. THE ROOT CAUSE OF THE ISSUE IS NOT KNOWN; HOWEVER, THE QC VARIATION AND DOWNWARD TREND WERE NOT OF SUFFICIENT MAGNITUDE TO CAUSE AN OIR LO (OUT OF INSTRUMENT RANGE, LOW) RESULT. THEREFORE, THE ISSUE APPEARS TO HAVE BEEN ISOLATED TO THE ONE PATIENT SAMPLE. (B)(4).
CUSTOMER CALLED TO REPORT THAT QUALITY CONTROL (QC) RESULTS FROM THE UNICEL DXC 600 SYNCHRON SYSTEM WERE TRENDING LOW, AND THAT ONE PATIENT SAMPLE RECOVERED WITH A SUPPRESSED RESULT, OIR LO (OUT OF INSTRUMENT RANGE, LOW), FOR AMMONIA. CUSTOMER STATED THAT NO ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY; THEREFORE, PATIENT TREATMENT WAS NOT AFFECTED. CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF A LABORATORY COAT AND GLOVES. THERE WAS NO DEATH, INJURY OR CHANGE TO PATIENT TREATMENT ATTRIBUTED TO OR ASSOCIATED WITH THIS COMPLAINT. THE FIELD SERVICE ENGINEER (FSE) FIRST INSPECTED THE INSTRUMENT FOR THE AMMONIA QC ISSUE ON (B)(4) 2012. NO REPORTS OF ANY ERRONEOUS PATIENT RESULTS ARE ASSOCIATED WITH THIS ISSUE. THE FSE REPLACED THE DEGASSER AND THE CARTRIDGE CHEMISTRY (CC) SAMPLE PROBE. THE FSE RETURNED ON (B)(4) 2012, AND NOTED THAT SINCE THE REAGENT PROBES AND DEGASSER WERE REPLACED, INSTRUMENT PRECISION HAD FAVORABLY INCREASED. UPON THIS VISIT, IT WAS SUSPECTED THAT THE AMMONIA QC ERROR OCCURRED AS A RESULT OF A PREANALYTICAL ISSUE (POSSIBLE FIBRIN).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 600 SYNCHRON SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER, INC. | DXC 600 PRO |