FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 24750931 · Received April 1, 2026

Report

Report Number
2955842-2026-20030
Event Type
Malfunction
Date Received
April 1, 2026
Date of Event
March 6, 2026
Report Date
May 29, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110720
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED COMPLAINT. THE FSE WAS NOT ABLE TO REPRODUCE THE ISSUE; HOWEVER, THE FSE REPLACED THE UNIVERSAL SURGICAL MANIPULATOR. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. AS OF THE DATE OF THIS REPORT, THE USM HAS NOT YET BEEN RECEIVED BY INTUITIVE SURGICAL, INC. (ISI) FOR EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED VENTRAL HERNIA TAPP SURGICAL PROCEDURE, THE USER INFORMED AN INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) THAT ARM 4 WAS NOT RECOGNIZING INSTRUMENTS. TROUBLESHOOTING STEPS INCLUDED CONFIRMING THAT ARM 4 WOULD NOT ENGAGE INSTRUMENTS, ALTHOUGH IT WAS UNDETERMINED IF THIS ISSUE AFFECTED ONLY ONE INSTRUMENT OR MULTIPLE INSTRUMENTS. LOGS CONFIRMED THE ENGAGEMENT FAILURE, INDICATING A POSSIBLE FAILURE WITH ARM 4 OR THE INSTRUMENT/ADAPTER. THE USER CONTINUED WITH THE PROCEDURE WITHOUT FURTHER ISSUES. NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED. ISI FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED ADDITIONAL INFORMATION: THE REPORTER CONFIRMED THAT THE SURGEON WAS ABLE TO COMPLETE THE SURGERY BY MAKING ARM 4 THE NON-WORKING ARM SO THERE WAS NO DELAY IN CARE. THE PROBLEM WAS QUICKLY REMEDIED WHEN THE FIELD SERVICE ENGINEER CAME TO EXCHANGE THE ARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
304086 DAVINCI XI PATIENT SIDE CART, 4-ARM NAY INTUITIVE SURGICAL, INC 380652-50 N/A 00886874110720

Patients

Seq Age Sex Outcome Treatment
1