UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM
Report
- Report Number
- 2050012-2012-00597
- Event Type
- Malfunction
- Date Received
- March 1, 2012
- Date of Event
- February 8, 2012
- Report Date
- February 8, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- CGZ
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE SAMPLES WERE TESTED IN 16X100 OPEN TUBES. QC WAS WITHIN LAB-ESTABLISHED RANGES PRIOR TO THE EVENT , BUT WAS LOW AFTER THE EVENT. SERVICE WAS DISPATCHED FOR THIS EVENT. THE FSE FOUND BLUE-GREY BUILD UP AROUND THE CO2 ELECTRODES AND A WHITE BUILDUP IN THE CL PORT. THERE WAS SOME BUILD UP IN THE LINE FROM THE INJECTION PORT TO THE WASTE SOLENOID VALVE. EIC (ELECTROLYTE INJECTION CUP) WAS CLEANED OUT. THE FSE REPLACED SODIUM (NA) REFERENCE ELECTRODE, NA MEASURE ELECTRODE, AND THE CARBON BRIDGE. CONTROLS WERE RUN AND ALL RESULTS WERE ON THE ESTABLISHED MEANS. THE INSTRUMENT PERFORMANCE WAS VERIFIED. AS OF (B)(4) 2012, THE CUSTOMER HAD NOT SEEN ANY CL ISSUES IN THE LAST TWO FULL NIGHTS OF OPERATION.
A CUSTOMER REPORTED TO BECKMAN COULTER, INC (BEC) THAT CHLORIDE (CL) RESULTS GENERATED BY UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEM HAD BEEN DRIFTING LOW AFTER APPROXIMATELY 200 - 300 SAMPLES WERE RUN. NO ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. THE CUSTOMER REPORTED THAT RECALIBRATING THE ASSAY BROUGHT THE QC AND PATIENT RESULTS BACK INTO THE EXPECTED RANGES. THE CUSTOMER VERBALLY PROVIDED A FEW EXAMPLES, OF WHICH THE LARGEST DIFFERENCE WAS 10 MMOL/L BETWEEN THE INITIAL AND REPEAT RESULTS. THE INITIAL AND THE REPEAT RESULTS AFTER RECALIBRATION WERE 92 AND 102 MMOL/L, RESPECTIVELY, AND THIS DIFFERENCE EXCEEDS ASSAY PRECISION CLAIMS. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | CGZ | BECKMAN COULTER, INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |