THERMO SCIENTIFIC
Report
- Report Number
- MW5024444
- Event Type
- Malfunction
- Date Received
- February 27, 2012
- Date of Event
- December 28, 2011
- Report Date
- February 8, 2012
- Manufacturer
- THERMO FISHER SCIENTIFIC
- Product Code
- LXG
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- MI, US
- Reporter Occupation
- PHARMACIST
Narratives
ON (B)(6) 2011 AT APPROXIMATELY 1030, ONE OF THE PHARMACY TECHNICIANS NOTICED A "BURNING SMELL" THAT WAS ORIGINATING FROM THE BIOLOGICAL SAFETY CABINET. THE TECHNICIAN WAS IMMEDIATELY INSTRUCTED TO TURN OFF THE HOOD. A CERTIFIED THERMO FISHER SCIENTIFIC SERVICE PERSON REMOVED THE COVER-PLATE ON THE BIOLOGICAL SAFETY CABINET AND NOTED THAT THE COMPUTERIZED MOTOR SPEED CONTROL BOARD HAD BURNT. HE ALSO COMMENTED THAT THE POWER SUPPLY WIRE TO THE ELECTRICAL MOTOR FROM THE CONTROL BOARD HAD REACHED SUCH A HIGH TEMPERATURE THAT IT HAD DISINTEGRATED WHEN HE REMOVED IT. THE WIRING DIAGRAMS INDICATED 16-18 GAUGE WIRING IN THE UNIT. THE SERVICE TECHNICIAN NOTED THAT THE BIOLOGICAL SAFETY CABINET WAS USING ALMOST 13.5 AMPS. THE MOTOR SPEED CONTROL BOARD WAS REPLACED AND THE ELECTRICAL REQUIREMENTS DECREASED. THERE HAVE BEEN OTHER REPAIR ISSUES WITH THIS BIOLOGICAL SAFETY CABINET. THIS IS DEFINITELY A QA ISSUE THAT COULD HAVE LED TO A CATASTROPHIC FIRE IN AN ONCOLOGY FACILITY. THIS PARTICULAR BIOLOGICAL SAFETY CABINET IS UTILIZED TO COMPOUND MANY OF THE PHASE I DRUGS IN OUR ONCOLOGY PTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMO SCIENTIFIC | THERMO ELECTRON FORMA CLASS II, A2 BIOLOGICAL SAFETY CABINET | LXG | THERMO FISHER SCIENTIFIC | 1286 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |