LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2026-03097
- Event Type
- Injury
- Date Received
- March 31, 2026
- Date of Event
- March 19, 2026
- Report Date
- March 31, 2026
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- UDI-DI
- 00855778005005
- PMA / PMN Number
- P010030
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THE MONITOR AND ELECTRODE BELT HAVE NOT BEEN RETURNED FOR EVALUATION. BASED ON THE DATA AVAILABLE AT THIS TIME, THERE IS NO INDICATION OF A DEVICE MALFUNCTION CAUSING OR CONTRIBUTING TO THE INAPPROPRIATE TREATMENT. INAPPROPRIATE DEFIBRILLATIONS ARE AN ANTICIPATED RISK ASSOCIATED WITH THE USE OF THE LIFEVEST. PATIENTS ARE INSTRUCTED THROUGH ALARMS, VOICE MESSAGES, IFU, AND TRAINING TO PRESS THE RESPONSE BUTTONS TO PREVENT AN INAPPROPRIATE DEFIBRILLATION. THE CURRENT COMMERCIAL INAPPROPRIATE DEFIBRILLATION RATE IS CONSISTENT WITH THE OBSERVED RATE DURING THE PIVOTAL CLINICAL TRIAL G960083 (0.69% PER PATIENT-MONTH WITH 90% CONFIDENCE). A SUMMARY OF THE SAFETY AND EFFECTIVENESS DATA (SSED), INCLUDING THE INAPPROPRIATE DEFIBRILLATION SAFETY OBJECTIVE SUPPORTING FDA'S APPROVAL OF THE LIFEVEST, CAN BE FOUND AT HTTP://WWW.ACCESSDATA.FDA.GOV/CDRH_DOCS/PDF/P010030B.PDF.
A PATIENT RECEIVED AN INAPPROPRIATE SHOCK. AF WITH RVR CONTRIBUTED TO THE FALSE DETECTION. THE RESPONSE BUTTONS WERE NOT PRESSED DURING THE EVENT. THE PATIENT CONTINUES WEARING LIFEVEST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 799463 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | LIFEVEST WCD 4000 SYSTEM | 00855778005005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |