FDA Adverse Event Other Summary report: N

ROTATING CF RESECTOSCOPE INNER SHEATH

MDR report key: 2474225 · Received February 28, 2012

Report

Report Number
1519132-2012-00014
Event Type
Other
Date Received
February 28, 2012
Date of Event
January 27, 2012
Report Date
February 2, 2012
Manufacturer
GYRUS ACMI, INC.
Product Code
FDC
PMA / PMN Number
PREAMEND
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSP OF THE INSTRUMENT CONFIRMS THAT THE CERAMIC TIP IS BROKEN OFF THE SHEATH TUBE. THE BROKEN PIECE OF CERAMIC WAS RETURNED WITH THE UNIT. A BURN IS NOTED ON THE BROKEN PIECE OF CERAMIC. THE BALANCE OF THE CERAMIC TIP REMAINS SECURELY GLUED INSIDE THE DISTAL END OF THE SHEATH TUBE. ALSO NOTED ARE MULTIPLE DENTS ALONG THE STEEL TUBE. THE EXACT CAUSE OF THE BREAKAGE OF THE CERAMIC TIP CANNOT BE DETERMINED. DUE TO THE PRESENCE OF THE BURN IN THE PIECE OF THE CERAMIC TIP THAT WAS RETURNED AND THE DENTS/DAMAGE ON THE STEEL TUBE AND THE FACT THAT THE ADHESIVE ON THE BASE OF THE TIP HAS RETAINED THE PROXIMAL PORTION IN THE TUBE, THE CAUSE OF THIS FAILURE IS DETERMINED TO BE CAUSED BY THE CUSTOMER.

Description of Event or Problem · 1

DURING A SURGICAL PROCEDURE THE CERAMIC TIP OF THE RESECTOSCOPE INNER SHEATH FRACTURED AND BROKE INSIDE THE PT. ALL PIECES WERE RETRIEVED. NO PT HARM WAS REPORTED.