ROTATING CF RESECTOSCOPE INNER SHEATH
Report
- Report Number
- 1519132-2012-00014
- Event Type
- Other
- Date Received
- February 28, 2012
- Date of Event
- January 27, 2012
- Report Date
- February 2, 2012
- Manufacturer
- GYRUS ACMI, INC.
- Product Code
- FDC
- PMA / PMN Number
- PREAMEND
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
VISUAL INSP OF THE INSTRUMENT CONFIRMS THAT THE CERAMIC TIP IS BROKEN OFF THE SHEATH TUBE. THE BROKEN PIECE OF CERAMIC WAS RETURNED WITH THE UNIT. A BURN IS NOTED ON THE BROKEN PIECE OF CERAMIC. THE BALANCE OF THE CERAMIC TIP REMAINS SECURELY GLUED INSIDE THE DISTAL END OF THE SHEATH TUBE. ALSO NOTED ARE MULTIPLE DENTS ALONG THE STEEL TUBE. THE EXACT CAUSE OF THE BREAKAGE OF THE CERAMIC TIP CANNOT BE DETERMINED. DUE TO THE PRESENCE OF THE BURN IN THE PIECE OF THE CERAMIC TIP THAT WAS RETURNED AND THE DENTS/DAMAGE ON THE STEEL TUBE AND THE FACT THAT THE ADHESIVE ON THE BASE OF THE TIP HAS RETAINED THE PROXIMAL PORTION IN THE TUBE, THE CAUSE OF THIS FAILURE IS DETERMINED TO BE CAUSED BY THE CUSTOMER.
DURING A SURGICAL PROCEDURE THE CERAMIC TIP OF THE RESECTOSCOPE INNER SHEATH FRACTURED AND BROKE INSIDE THE PT. ALL PIECES WERE RETRIEVED. NO PT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ROTATING CF RESECTOSCOPE INNER SHEATH | FESECTOSCOPE INNER SHEATH | FDC | GYRUS ACMI, INC. | ERIS-CF25 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |