FDA Adverse Event
Malfunction
Summary report: N
MEDIPURE
MDR report key: 2474156
·
Received February 23, 2012
Report
- Report Number
- 2474156
- Event Type
- Malfunction
- Date Received
- February 23, 2012
- Date of Event
- February 7, 2012
- Report Date
- February 22, 2012
- Manufacturer
- PRAXAIR, INC.
- Product Code
- CAN
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- IA, US
Narratives
Description of Event or Problem · 1
CYLINDER WAS PAINTED WITH THE INDUSTRY STANDARD COLOR OF BLUE (NITROUS OXIDE) INSTEAD OF GREEN (OXYGEN). PAPER LABEL FOR CONTENTS OF THE CYLINDER LABELED AS OXYGEN. VALVE STEM ON THE CYLINDER WAS PIN INDEXED FOR THE INDUSTRY STANDARD TO ALLOW AN OXYGEN REGULATOR.======================MANUFACTURER RESPONSE FOR MEDIPURE MEDICAL GAS CYLINDER (PER SITE REPORTER).======================MANUFACTURER WILL CONDUCT THE ANALYSIS OF THE CONTENTS OF THE CYLINDER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEDIPURE | MEDICAL GAS CYLINDER | CAN | PRAXAIR, INC. | UN1072 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |