FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE

MDR report key: 24741386 · Received March 31, 2026

Report

Report Number
9610773-2026-02535
Event Type
Malfunction
Date Received
March 31, 2026
Date of Event
February 12, 2026
Report Date
March 31, 2026
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761074971
PMA / PMN Number
K190744
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE LLK (LASER LIGHT CABLE) PLUG OF THE VIDEO CABLE SECTION HAD FOREIGN MATERIAL CAUSE COULD NOT BE ESTABLISHED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GYNECOLOGIC LAPAROSCOPE HAD LLK (LASER LIGHT CABLE) PLUG OF THE VIDEO CABLE SECTION HAD FOREIGN MATERIAL. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
806978 VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE LAPAROSCOPE, GYNECOLOGIC HET OLYMPUS WINTER & IBE GMBH WA50042A 04042761074971

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown