ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT
Report
- Report Number
- 3005094123-2026-00148
- Event Type
- Malfunction
- Date Received
- March 30, 2026
- Date of Event
- March 7, 2026
- Report Date
- May 12, 2026
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- MMI
- UDI-DI
- 00380740160630
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GG
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SECTION A1 PATIENT IDENTIFIER COMPLETE INFORMATION: (B)(6). THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 8P13 (ALINITY I STAT HIGH SENSITIVE TROPONIN-I) THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 4Z21 (ALINITY I STAT HIGH SENSITIVITY TROPONIN-I) WITH 510K/PMA/BLA NUMBER K202525. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.
THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULT GENERATED ON THE ALINITY I PROCESSING MODULE FOR A 49-YEAR-OLD MALE PATIENT BROUGHT INTO THE EMERGENCY DEPARTMENT PRESENTING WITH CHEST PAIN. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6). FIRST RESULT = 227.1 NG/L. BASED ON THIS ELEVATED RESULT, THE PATIENT UNDERWENT A CORONARY ANGIOGRAM. SIX HOURS LATER, A REPEAT TEST WAS PERFORMED, AND THE SAMPLE GENERATED A RESULT OF 1.1 NG/L. THE FIRST RESULT WAS QUESTIONED DUE TO THE SIGNIFICANT DIFFERENCE. A THIRD SAMPLE WAS TESTED FOR TROPONIN-I AND GENERATED A RESULT OF 2.1 NG/L. THE ORIGINAL SAMPLE WAS RETESTED AND GENERATED A RESULT OF 0.2 NG/L. THERE WAS NO FURTHER IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 264823 | ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT | IMMUNOASSAY METHOD, TROPONIN SUBUNIT | MMI | ABBOTT IRELAND DIAGNOSTICS DIVISION | 81564UD00 | 00380740160630 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Male | ALNTY I PROCESSING MODU, 03R65-01, (B)(6) |