FDA Adverse Event Malfunction Summary report: N

COBAS B 221 <6> SYSTEM

MDR report key: 24728310 · Received March 30, 2026

Report

Report Number
1823260-2026-01157
Event Type
Malfunction
Date Received
March 30, 2026
Date of Event
March 11, 2026
Report Date
May 11, 2026
Manufacturer
ROCHE DIAGNOSTICS
Product Code
CHL
UDI-DI
04015630018345
PMA / PMN Number
K032311
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TC
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION DETERMINED THE ISSUE IS CONSISTENT WITH INSUFFICIENT PRE-ANALYTIC SAMPLE HANDLING. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.

Additional Manufacturer Narrative · 0

ELECTRODE LOT NUMBERS AND EXPIRATION DATES WERE REQUESTED, BUT NOT PROVIDED. THE INVESTIGATION IS ONGOING. MEDWATCH FIELD D2B PROCODE: PRODUCT CODE CHL APPLIES TO THE PCO2, PO2, AND PH PARAMETERS. OTHER PRODUCT CODES ASSOCIATED WITH THE COBAS B 221 <6> SYSTEM ARE CEM, CDS, CGA, GGF, GGZ, GHS, GKR, JFP, JGS, KHG, AND KHP.

Description of Event or Problem · 0

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT PO2, PCO2, CHCO3, BE, AND PH RESULTS FOR ONE PATIENT SAMPLE TESTED ON A COBAS B 221 <6> SYSTEM. THE SAMPLE WAS INITIALLY TESTED AND REPEATED ON (B)(6) 2026, RESULTING IN THE FOLLOWING VALUES: PO2 = 159.3 MMHG WITH A DATA FLAG FOR THE FIRST VALUE AND 99.8 MMHG FOR THE REPEAT VALUE PCO2 = 16.1 MMHG WITH A DATA FLAG FOR THE FIRST VALUE AND 30.2 MMHG WITH A DATA FLAG FOR THE REPEAT VALUE. CHCO3 = 12.9 MMOL/L FOR THE FIRST VALUE AND 20.6 MMOL/L FOR THE REPEAT VALUE . BE = - 6.6 MMOL/L FOR THE FIRST VALUE AND - 2.1 MMOL/L FOR THE REPEAT VALUE. PH = 7.522 WITH A DATA FLAG FOR THE FIRST VALUE AND 7.452 WITH A DATA FLAG FOR THE REPEAT VALUE. TESTING FOR THE SAMPLE WAS REPEATED ON (B)(6) 2026 WITH A SHORTER DELAY BETWEEN FIRST AND REPEAT MEASUREMENTS. THE FOLLOWING VALUES WERE OBTAINED: PO2 = 162.1 MMHG WITH A DATA FLAG FOR THE FIRST VALUE AND 156.4 MMHG WITH A DATA FLAG FOR THE REPEAT VALUE. PCO2 = 39.0 MMHG FOR THE FIRST VALUE AND 38.8 MMHG FOR THE REPEAT VALUE. CHCO3 = 19.3 MMOL/L FOR THE FIRST VALUE AND 19.0 MMOL/L FOR THE REPEAT VALUE . BE = - 6.2 MMOL/L FOR THE FIRST VALUE AND - 6.7 MMOL/L FOR THE REPEAT VALUE. PH = 7.313 WITH A DATA FLAG FOR THE FIRST VALUE AND 7.308 WITH A DATA FLAG FOR THE REPEAT VALUE.