FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE

MDR report key: 24723802 · Received March 30, 2026

Report

Report Number
9610773-2026-02441
Event Type
Malfunction
Date Received
March 30, 2026
Report Date
March 29, 2026
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761074971
PMA / PMN Number
K190744
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

B3- DATE OF EVENT IS UNKNOWN. ADDITIONAL INFORMATION WILL BE PROVIDED ONCE AVAILABLE. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, CAUSE NOT ESTABLISHED FOR LLK (LASER LIGHT CABLE PLUG OF THE VIDEO CABLE SECTION CONTAINS FOREIGN MATERIAL, AND CAUSE COULD BE COMPONENT FAILURE FOR FIBER BONDING IN THE OUTER TUBE AREA (DISTAL END) IS DAMAGED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING DEVICE EVALUATION THE LAPAROSCOPE, GYNECOLOGIC HAD DAMAGED FIBER BONDING IN THE OUTER TUBE AREA (DISTAL END), AND THE LLK (LASER LIGHT CABLE) PLUG OF THE VIDEO CABLE SECTION CONTAINED FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
787434 VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE LAPAROSCOPE, GYNECOLOGIC HET OLYMPUS WINTER & IBE GMBH WA50042A 04042761074971

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown