POWERED WHEELCHAIR
Report
- Report Number
- 1525712-2012-00298
- Event Type
- Malfunction
- Date Received
- February 29, 2012
- Report Date
- April 11, 2012
- Manufacturer
- INVACARE TAYLOR STREET
- Product Code
- ITI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PATIENT
Narratives
NO RMA HAS BEEN INITIATED FOR THIS ISSUE. MODEL TDXSP-CG, (B)(4) IS APPROXIMATELY 1 YEAR 8 MONTHS OLD. THE OWNER'S MANUAL PART NUMBER 1143190 REV I (JUN-10) WAS ISSUED WITH THIS DEVICE. THE OWNER'S MANUAL IS ALSO FOUND ON-LINE AT INVACARE.COM. IT IS UNKNOWN IF THE CONSUMER HAS FULLY READ AND UNDERSTANDS THE OWNER'S MANUAL. DOCUMENTATION PROVIDES WARNINGS, CAUTIONS, AND INSTRUCTIONS FOR SAFELY USING THE DEVICE. IF THE CONSUMER DOES NOT UNDERSTAND THE WRITTEN WARNINGS, CAUTIONS OR INSTRUCTIONS THEN THEY SHOULD CONTACT INVACARE. THE CONSUMER IS A MALE WHOSE AGE, HEIGHT AND WEIGHT ARE UNKNOWN. THE CONSUMERS MEDICAL CONDITION, STABILITY AND MEDICATION REGIMEN ARE UNKNOWN. THE CONSUMERS TECHNIQUE WHILE USING THE DEVICE IS UNKNOWN. THE MAINTENANCE HISTORY OF THE DEVICE IS UNKNOWN.
INITIAL (B)(4) ISSUED MFG REPORT # 1525712-2012-00298. THE COMPONENTS, MOTOR MODEL #11416888, SERIAL NUMBER (B)(4), DATE CODE (B)(4) AND CONTROLLER MODEL #1136898 REV F, SERIAL # (B)(4) HAS BEEN RETURNED FOR AN EVALUATION. THE GEARBOX SHAFT HAD A SIGNIFICANT AMOUNT OF HAIR WRAPPED AROUND IT. THE MOTOR WAS NOISY WHEN FUNCTIONALLY TESTED BUT NO OTHER DISCREPANCIES WERE OBSERVED. THE CONTROLLER WAS ALSO TESTED AND NO DISCREPANCIES WERE OBSERVED. NO RMA HAS BEEN INITIATED FOR THIS ISSUE. MODEL TDXSP-CG, SERIAL NUMBER/DATE CODE (B)(4) IS APPROXIMATELY 1 YEAR 8 MONTHS OLD. THE OWNER'S MANUAL PART NUMBER 1143190 REV I (JUN-10) WAS ISSUED WITH THIS DEVICE. THE OWNER'S MANUAL IS ALSO FOUND ON-LINE AT INVACARE.COM. IT IS UNKNOWN IF THE CONSUMER HAS FULLY READ AND UNDERSTANDS THE OWNER'S MANUAL. DOCUMENTATION PROVIDES WARNINGS, CAUTIONS, AND INSTRUCTIONS FOR SAFELY USING THE DEVICE. IF THE CONSUMER DOES NOT UNDERSTAND THE WRITTEN WARNINGS, CAUTIONS OR INSTRUCTIONS THEN THEY SHOULD CONTACT INVACARE. THE CONSUMER IS A MALE WHOSE AGE, HEIGHT AND WEIGHT ARE UNKNOWN. THE CONSUMERS MEDICAL CONDITION, STABILITY AND MEDICATION REGIMEN ARE UNKNOWN. THE CONSUMERS TECHNIQUE WHILE USING THE DEVICE IS UNKNOWN. THE MAINTENANCE HISTORY OF THE DEVICE IS UNKNOWN.
CONSUMER ALLEGES THAT THE TILT ACTUATOR STARTED TILTING BACK WHILE HE WAS UNDERNEATH THE STEERING WHEEL OF HIS VAN; HE WAS NOT DRIVING AT THAT TIME. MINOR INJURY ALLEGED.
CONSUMER ALLEGES THAT THE TILT ACTUATOR STARTED TILTING BACK WHILE HE WAS UNDERNEATH THE STEERING WHEEL OF HIS VAN; HE WAS NOT DRIVING AT THAT TIME. MINOR INJURY ALLEGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | POWERED WHEELCHAIR | 890.3860 | ITI | INVACARE TAYLOR STREET | TDXSP-CG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |