FDA Adverse Event Malfunction Summary report: N

3M ATTEST STEAM CHEMICAL INTEGRATOR WITH EXTENDER

MDR report key: 24709562 · Received March 30, 2026

Report

Report Number
MW5186046
Event Type
Malfunction
Date Received
March 30, 2026
Date of Event
March 3, 2026
Report Date
March 24, 2026
Manufacturer
3M/SOLVENTUM US LLC
Product Code
JOJ
Product Problem
Yes
Report Source
Voluntary report
Reporter Occupation
RISK MANAGER
Health Professional
Yes

Narratives

Description of Event or Problem · 0

AFTER STEAM STERILIZATION LOAD COMPLETED, SPD TECH HAD IDENTIFIED MULTIPLE INSTRUMENTS IN PEEL POUCHES THAT HAD A CHEMICAL INTEGRATOR (CI) WITH INK LEAKING FROM WITHIN. SPD MANAGER RULED OUT STERILIZER EQUIPMENT FAILURE BY VERIFYING THAT THE PEEL PACKS WITH DAMAGED CI'S WERE RUN IN DIFFERENT AUTOCLAVES AND CYCLES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
770152 3M ATTEST STEAM CHEMICAL INTEGRATOR WITH EXTENDER INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS JOJ 3M/SOLVENTUM US LLC 1243RE SY012029

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown