FDA Adverse Event
Malfunction
Summary report: N
3M ATTEST STEAM CHEMICAL INTEGRATOR WITH EXTENDER
MDR report key: 24709562
·
Received March 30, 2026
Report
- Report Number
- MW5186046
- Event Type
- Malfunction
- Date Received
- March 30, 2026
- Date of Event
- March 3, 2026
- Report Date
- March 24, 2026
- Manufacturer
- 3M/SOLVENTUM US LLC
- Product Code
- JOJ
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Occupation
- RISK MANAGER
- Health Professional
- Yes
Narratives
Description of Event or Problem · 0
AFTER STEAM STERILIZATION LOAD COMPLETED, SPD TECH HAD IDENTIFIED MULTIPLE INSTRUMENTS IN PEEL POUCHES THAT HAD A CHEMICAL INTEGRATOR (CI) WITH INK LEAKING FROM WITHIN. SPD MANAGER RULED OUT STERILIZER EQUIPMENT FAILURE BY VERIFYING THAT THE PEEL PACKS WITH DAMAGED CI'S WERE RUN IN DIFFERENT AUTOCLAVES AND CYCLES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 770152 | 3M ATTEST STEAM CHEMICAL INTEGRATOR WITH EXTENDER | INDICATOR, PHYSICAL/CHEMICAL STERILIZATION PROCESS | JOJ | 3M/SOLVENTUM US LLC | 1243RE | SY012029 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |