FDA Adverse Event Malfunction Summary report: N

DAVINCI 5

MDR report key: 24708395 · Received March 26, 2026

Report

Report Number
2955842-2026-18881
Event Type
Malfunction
Date Received
March 26, 2026
Date of Event
October 23, 2025
Report Date
March 26, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874119662
PMA / PMN Number
K232610
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. BASED ON THE FIELD EVALUATION, THIS REPORTED EVENT WAS CONFIRMED. CUSTOMER REPORTED ERROR 297 AT START UP. FSE REPROGRAMMED SYSTEM'S COMMON COMPUTE CONTROLLER (CCC) TO RESOLVE THE ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. THE COMPLAINT WAS CONFIRMED BASED ON THE FIELD EVALUATION. THE PROBABLE ROOT CAUSE IS DUE TO THE SOFTWARE CONVERTING TO A NON-CLINICAL (GOLDEN IMAGE). THIS ISSUE CAN BE RESOLVED BY HAVING THE FSE REPROGRAM THE SYSTEM. THIS ISSUE IS ALSO CAPTURED IN THE CLINICAL RISK ANALYSIS, GEN5, SYSTEM (818555-10, REV V) VIA RISK ID G5-SYS-5637.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER CONTACTED THE TECHNICAL SUPPORT ENGINEER (TSE) AND REPORTED ERROR 297 AT STARTUP. THE SYSTEM IS NOT POSTING TO ONSITE. POPUP LOGS SHOW 297 AND 307 WITH SOME 311'S. THE 311 ERROR IS A GOLDEN IMAGE ERRORS. THE REPORTING NODE FOR 311 IS SURGEON SIDE CONSOLE(SSC) NODE 33. MOST LIKELY REPROGRAMING WILL RESOLVE THE ISSUE BUT BRING A COMMON COMPUTER CONTROLLER (CCC) AND SIOX CARD CAGE FOR SIM. THIS IS A DOWN SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
18708 DAVINCI 5 VISION SIDE CART NAY INTUITIVE SURGICAL, INC 380746-40 N/A 00886874119662

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES